KVO and Arming Injector

Independent KVO (Keep Vein Open)

1. Ensure that no air is in the syringe or fluid path
2. Press the Air Confirmation button on injector head or acknowledge

on the DCU that air has been removed

3. Touch the Start button to begin KVO

KVO delivers a pulse of saline to ensure line patency

To ARM if Independent KVO Has Been Used

1. If KVO has been started, touch the ARM button on display screen 
2.  Press hand switch to begin the injection

To ARM Injector Without Using KVO

1. Ensure that no air is in the syringe or fluid path
2. Press the Air Confirmation button on injector head or 

confirm on the DCU that air has been removed

3. Touch ARM on Display Screen
4. Press the hand switch to begin the injection

7

Medrad® Spectris Solaris® EP QuickTips

NOTE: KVO can be started at any time after
arming the injector. To start KVO, touch
Start button before the injection is started.

Содержание Medrad Spectris Solaris EP

Страница 1: ...QuickTips...

Страница 2: ......

Страница 3: ...Mobile Manual 205580 Refer to the Operations Manual for compete instructions and user warnings Powering Up and Down Power UP 1 To turn on lift the clear plastic cover on the back of the Display Contr...

Страница 4: ...ge locks into place 4 Advance the plungers completely forward by pressing the Enable button A then pressing both Forward Load Switches B The pistons will stop when fully advanced 5 Remove protective c...

Страница 5: ...ath Prime T connector and tubing by advancing plungers with the forward or reverse switches or the manual piston control knobs Advance contrast up to the bottom of the check valve then prime the rest...

Страница 6: ...ppropriate phase type 3 Enter in the appropiate values Flow Rate 1 Touch the corresponding box 2 Enter the desired flow rate using the keypad 3 Touch enter to confirm Volume 1 Touch the corresponding...

Страница 7: ...Programming continued Pressure Limit 1 Touch Set on Pressure Limit field 2 From drop down list select the appropriate psi from 100 325 6 Medrad Spectris Solaris EP QuickTips...

Страница 8: ...sed 1 If KVO has been started touch the ARM button on display screen 2 Press hand switch to begin the injection To ARM Injector Without Using KVO 1 Ensure that no air is in the syringe or fluid path 2...

Страница 9: ...lay desired protocol on screen by setting Flow Rates Volumes and delay rates if applicable 2 Touch Store button 3 Type in desired protocol name using the keyboard 4 Touch enter Protocol has now been s...

Страница 10: ...otocol will be displayed on main screen 3 The pressure limit is changed to the stored protocol pressure limit value Injection History 1 From the Main protocol screen select HISTORY button 2 The screen...

Страница 11: ...g the iCBC from the wall outlet or remove the power plug at the iCBC box 4 Insert the battery into the Scan Room Unit 5 Plug in iCBC 6 Plug in DCU and switch it on Ensure iCBC icon on DCU has yellow b...

Страница 12: ...or reproduced in any form mechanical or electronic including photocopying tape recording microcopying or stored in a data carrier or computer system without written consent of Bayer HealthCare Bayer...

Отзывы: