TESTING
24
®
Disinfecting Your Meter
Supplies Needed for Disinfecting:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
• Timing device
1.
Before disinfecting, clean the meter as described in
Cleaning Your Meter
.
Cleaning and Disinfection
For proper disinfection, you must keep all meter
surfaces wet for 60 seconds.
Using a new germicidal wipe, carefully wipe all outer
surfaces of your meter until wet.
Do not allow cleaning
solution to run into the meter through areas such as around
the buttons or the meter’s test strip or data ports.
2.
Dry all meter surfaces and test strip port using a clean
paper towel if needed.
Cleaning Your Lancing Device
provided by Bayer
Supplies Needed for Cleaning:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
1.
Clean the lancing device provided by Bayer with
germicidal wipes until all soil is removed.
2.
Dry as necessary with a clean paper towel.
Disinfecting Your Lancing Device
provided by Bayer
Supplies Needed for Disinfecting:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
• Timing device