![Bayer HealthCare CONTOUR NEXT LINK Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/bayer-healthcare/contour-next-link/contour-next-link_user-manual_2712803020.webp)
TESTING
32
TESTING
33
OK
Result Under 20 mg/dL
Follow Medical Advice
Immediately.
E23
OK
Result Over 600 mg/dL
Wash hands or test
site. Repeat test with
new strip.
E21
OK
Over 600 mg/dL, Again
Follow Medical Advice
Immediately.
E22
Disinfecting Your Lancing Device (cont’d)
3.
Using a new Clorox
®
Germicidal Wipe, carefully
wipe all external and internal surfaces of the endcap
until wet.
4.
Use Clorox
®
Germicidal Wipes to disinfect all
exposed surfaces of your lancing device until wet.
5.
Dry all lancing device
surfaces and endcap
with a clean paper
towel if needed.
Measuring Range
Your meter’s measuring range is 20–600 mg/dL.
• If your blood sugar
reading is under
20 mg/dL, the
Follow Medical Advice
Immediately
screen will
display and the meter
will beep twice. Contact
your health care
professional.
• If your blood sugar
reading is over
600 mg/dL the next
screen will tell you to
retest. The meter will
beep twice.
• If results are still
over 600 mg/dL,
Follow Medical
Advice Immediately
is displayed.
Measuring Range
Disinfecting Y
our Lancing Device
For proper disinfection, you must keep all
lancing device surfaces wet for 60 seconds.
Signs of Deterioration:
Your M
ICROLET
®
2 lancing
device has been tested for 260 cycles of the cleaning
and disinfection procedure (one cycle per week for
5 years). The M
ICROLET
2 lancing device has been
demonstrated to withstand 5 years of cleaning and
disinfection without damage. However, if your device
malfunctions for any cause, contact Bayer Diabetes
Care Customer Service at 1-800-348-8100 (available
24 hours a day, 7 days a week) to troubleshoot and
provide a solution.
Warranty:
Your M
ICROLET
2 lancing device is
warranted for 5 years.
CAUTION:
• If your blood sugar reading is under 50 mg/dL,
follow medical advice immediately.
• If your blood sugar reading is over 250 mg/dL,
wash and dry your hands well and repeat the test
with a new strip. If you get a similar result, call
your health care professional as soon as possible.
• Always consult your health care professional
before changing your medication based on
C
ONTOUR
®
N
EXT
LINK test results.