Technical Information
A repeatability study was conducted with the
C
ONTOUR
®
N
EXT
EZ blood glucose monitoring system
using five venous whole blood specimens with glucose
levels from 47 to 326 mg/dL. Multiple replicates (n=100
per lot) were tested using multiple C
ONTOUR
N
EXT
EZ
blood glucose meters and three lots of C
ONTOUR
®
N
EXT
blood glucose test strips. The following precision
results were obtained.
System repeatability results for C
ONTOUR
N
EXT
EZ
meter using C
ONTOUR
N
EXT
test strips
Mean, mg/dL
Pooled
Standard
Deviation
Coefficient of
Variation, %
47
0.8
1.7
84
1.1
1.3
139
2.1
1.5
202
2.6
1.3
326
5.0
1.5
Service Information
If you have a problem and none of the problem solving
steps in this user guide help, contact Bayer Diabetes
Care Customer Service at 1-800-348-8100 (available
24 hours a day, 7 days a week). We have trained
specialists to help you.
Important
•
Speak to a Bayer Diabetes Care Representative
before returning your meter for any reason. He/
she will give you the information needed to get
your problem handled correctly and efficiently.
•
Have your C
ONTOUR
®
N
EXT
EZ blood glucose
meter and C
ONTOUR
®
N
EXT
test strips available
when you phone. It would also be helpful to have
a bottle of C
ONTOUR
®
N
EXT
control solution nearby.
Check List
This check list may be helpful when speaking with
Bayer Diabetes Care Customer Service:
1.
Locate the serial
number (A) and model
number (B) on the back
of the meter.
–––––––––––––––––––––
–––––––––––––––––––––
2.
Locate the C
ONTOUR
N
EXT
test strips’ expiration date
on the bottle.
–––––––––––––––––––––––––––––––
3.
Check the battery icon on the display.
(See page 47, Changing the Batteries.)
¯