17 - 89
17.7.3 Programmable Pressure System Performance
17.8 Power Cable Specifications
The specifications required by the Stellant with Certegra Workstation relative to the power cable (plug,
receptacle, and cord) are:
•
Operating Temperature: 60° C minimum
•
Receptacle Type: IEC-60320 C13
•
Normal Cord Voltage: 300 VAC minimum
•
Wire Gauge: 1.00 mm
2
minimum
•
Cord Type: IEC 60245-1, Annex A, Designation 53, or IEC 60227-1, Annex A, Designation 53,
Certified
•
Cord Length: 3 m maximum
The power cable must meet applicable plug, cord, and receptacle specifications including type, volt-
age, current, and safety approval markings for the country in which the power cable is being used.
17.9 Injection System Cybersecurity and IT Network Connection
17.9.1 System Cybersecurity Protection
System software cybersecurity controls include:
•
Internal data integrity checks
•
Remote access requires valid credentials and a user at the workstation to approve a remote
connection
•
Service access requires valid credentials
•
Access to the Operating System is restricted
•
Firewall and Anti-Virus
•
(Configurable) Masking Protected Health Information (PHI) when sharing with external sys-
tems
•
Network settings can only be configured by Bayer
•
Data Manager website access requires valid credentials
•
Certegra database encrypted and access restricted
Table 17 - 3: Programmable Pressure System Performance
Programmed Pressure Limit
High Pressure Disarm
325 psi (2241 kPa)
375 psi (2586 kPa)
300 psi (2068 kPa)
350 psi (2413 kPa)
250 psi (1724 kPa)
300 psi (2068 kPa)
225 psi (1551 kPa)
275 psi (1896 kPa)
200 psi (1379 kPa)
250 psi (1724 kPa)
150 psi (1034 kPa)
200 psi (1379 kPa)
100 psi (689 kPa)
150 psi (1034 kPa)
50 psi (345 kPa)
100 psi (689 kPa)
Содержание Certegra
Страница 2: ......
Страница 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Страница 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Страница 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Страница 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Страница 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Страница 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Страница 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Страница 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Страница 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Страница 103: ......