EN
USER’S MANUAL
ARD
36
3.1. DIVISION SETTING
Display shows “d x”, then press
key to select 1, 2, 5, 10, 20, 50.
Then press
key to confirm, and enter the decimal point setting.
3.2. DECIMAL POINT SETTING
Display shows “p x”, then press
key to select 0, 0.0, 0.00, 0.000, 0.0000, P.00000. Then press
key to confirm, and
enter the maximum capacity setting.
3.3. MAXIMUM CAPACITY SETTING
Display showsa “ FULL ” around 2-3 seconds then display shows “000.000”, the decimal point was set
as 3 decimal points, then press
key to move the flashing digits to the right in circle to select the digits that will increase the
number, then press
key to increase the number until the number appears, then repeat the above moves until display shows
the maximum capacity.
Then press
key to enter calibration setting.
3.4. ZERO POINT CALIBRATION
Display shows “CaL 0”, do not put any objects on the pan when display shows this, then scale will move to next calibration
setting.
3.5. SECOND POINT CALIBRATION
Display shows “00.000” (the decimal point was set as 3 digits)
Put on the weights on the pan which will be calibrate as second point then use
key to move the flashing digits to the right
in circle to select the digits that will increase the number, then press
key to increase the number until the number appears.
Then repeat the above moves until display shows the numbers which is equal to the weights put on the pan.
Press
key to confirm calibration finished.