07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
7-19
Troubleshooting
Troubleshooting Chart
Troubleshooting Chart
Table 7-4 identifies specific failures by symptoms rather than Failure
Identification Codes.
Table 7-4
Troubleshooting Chart
Symptom
Checks
Corrective Action
Power Issues
One or both pumps cannot be turned
on.
Check backup battery voltage in
accordance with “Automatic Test
Mode 1: Calibration Mode 1” on page
9-14.
Check front panel cables for short
circuits and proper connection.
Check CN601 and CN602 for proper
connection and continuity.
Sometimes this condition is followed
by a F49 and/or F94 code. Charge or
replace the battery to clear the
condition. Repair if necessary.
One or both pumps cannot be turned
off.
Check backup battery voltage in
accordance with “Automatic Test
Mode 1: Calibration Mode 1” on page
9-14.
Check front panel cables for short
circuits.
Check CN601 and CN602 for proper
connection and continuity.
Check if all 3 LEDs are on and the
constant alarm is on.
Sometimes this condition is followed
by a F49 and/or F94 code. Recharge
or replace the battery to clear the
condition. The battery compartment
may need to be opened and the battery
momentarily disconnected to turn off
the pump.
Disconnect then reconnect lithium
battery.
Replace front panel film.
Replace CPU board.
Replace Sensor board.
Pump does not operate long on battery
power.
Verify condition and capacity of
battery.
Verify continuity of the battery
harness.
Charge the battery uninterrupted for at
least 16 hours.
Replace the battery.
Repair or replace the battery harness.
Battery does not provide rated backup
power when AC power is not
available.
Perform “Battery Check” on page
Recharge or replace battery as
required.
Pressing a
START
key causes a
BATTERY LOW alarm.
Power the pump off and on again.
If the problem persists, recharge the
battery.
Repair as required.
If the problem persists, replace the
battery.
Содержание Flo-Gard 6301
Страница 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 20: ...1 General Information 1 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 30: ...2 Controls and Indicators 2 10 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 34: ...3 Administration Sets 3 4 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 36: ...4 Nurse Call 4 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 74: ...7 Troubleshooting Chart 7 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 110: ...8 Disassembly Reassembly 8 36 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 200: ...13 Overview 13 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 201: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 3 Figure 13 1 System Block Diagram ...
Страница 205: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 7 Figure 13 5 Power Supply Section ...
Страница 206: ...13 8 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 6 DC Power Supply ...
Страница 210: ...13 12 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 10 Master CPU Schematic ...
Страница 213: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 15 Figure 13 13 Address Decoder ...
Страница 216: ...13 18 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 16 Slave CPU Schematic ...
Страница 220: ...13 22 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 20 CPU PCB Sensor Block ...
Страница 221: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 23 Figure 13 21 PPI Block 1 ...
Страница 222: ...13 24 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 22 PPI Block 2 ...
Страница 224: ...13 26 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 24 Air Sensing Block ...
Страница 226: ...13 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 26 Display PCB ...
Страница 229: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 31 Figure 13 29 LCDs for Pump 1 ...
Страница 230: ...13 32 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 30 LCDs for Pump 2 ...
Страница 231: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 33 Figure 13 31 LED Lamps ...
Страница 232: ...13 34 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 32 Audible Alarm PCB ...
Страница 233: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 35 Figure 13 33 Backup Buzzer PCB ...
Страница 234: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 36 Figure 13 34 Terminal PCB ...
Страница 235: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 37 Figure 13 35 Accessories ...
Страница 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 251: ...07 19 B1 729 ...