Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C
Page 87
2.
If the
Bag currently on the load cell does not appear to be empty
message appears,
visually check the contents of the bag.
•
If the bag is empty, refer to The bag currently on the load cell does not appear to be
•
If the bag contains fluid:
a.
Tap
No
.
b.
At the
Operation Cancelled
message, tap
OK
.
c.
Remove the bag. Refer to Removing the Calibration Bag on Page 73.
d.
Attach an empty bag. Refer to Attaching the Patient Bag on Page 85.
e.
Return to Step 1.
Message
WARNING
If you tap
Yes
, the compounder will reset the measured weight to zero, despite
the fact that the bag contains fluid.
The finished solution may contain an unintended volume or ingredient, even if
the final measured weight is within the acceptable range. This unintended
volume or ingredient may result in patient harm.
The
Details
section of the MixCheck Report will indicate that you continued
compounding despite the warning that the bag did not appear to be empty.
For instructions on how to handle a finished bag in this situation, refer to your
facility’s protocol.
Содержание ExactaMix 2400 Compounder
Страница 2: ......
Страница 170: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 162 53007410 Rev C Sample MixCheck Report...
Страница 177: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 169 Sample Formula Report...
Страница 180: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder Page 172 53007410 Rev C Sample Log Report...
Страница 185: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 177 Sample Formulary Report...
Страница 199: ...Using Reports Operator Manual for the Baxter ExactaMix 2400 Compounder 53007410 Rev C Page 191 Sample Bag Usage Report...