4
Primary Infusions
4-32
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
10.
Press the
START
key to begin the infusion. COLLEAGUE
GUARDIAN infusions are indicated by the mortar and pestle icon
next to the channel on the Main Display screen.
A yellow triangle is displayed beside the label name if the drug
amount, diluent, or concentration was changed to deviate from the
standard COLLEAGUE GUARDIAN settings.
If a secondary infusion has been programmed but not started, a
pop-up appears (Figure 4-26) warning that a primary infusion is
about to start when a secondary infusion has been programmed. See
step 7 on page 4-22 for more information.
Programming a Primary COLLEAGUE GUARDIAN Infusion (Weight-Based)
1.
From the Main Display, access the Programming screen:
2.
From the Programming screen, press the
Colleague Guardian
soft
key.
The labels for which COLLEAGUE GUARDIAN limits have been
configured are displayed in a pop-up window (Figure 4-44). A
yellow triangle beside the label indicates that the standard
concentration can be edited if clinically necessary
3.
Use the
×Ø
keys (and
Page Up/Page Down
soft keys if necessary)
to highlight the desired label, then press the
Select
soft key.
Figure 4-43 Infusion with
COLLEAGUE GUARDIAN Limit
Override
Note:
Overriding
COLLEAGUE
GUARDIAN
Limits
If the clinical decision was to override the
COLLEAGUE GUARDIAN limits, the dose and
programming mode are displayed in red on a yellow
highlight (Figure 4-43) indicating that the programmed
dose is outside of the COLLEAGUE GUARDIAN limits.
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key or
the
Rate
or
Volume
keys.
press the desired
Channel
Select
key.
Figure 4-44 COLLEAGUE
GUARDIAN Label List
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...