Troubleshooting
Troubleshooting Failures
Channel Failures (Single Channel Pumps)
Figure 8-2 shows the Channel Failure alarm message on a single channel
pump. The procedure for responding to a channel failure is provided
below.
1.
Access the programming screen from the Main Display if necessary
by pressing the
Primary
or
Secondary
soft key. The parameters for
the infusion in progress when the channel failed are displayed (see
Figure 8-3 for an example). Record the parameters so the infusion
can be continued on another pump if necessary.
2.
Request a replacement pump immediately.
3.
Perform step Step 4 or Step 5 as appropriate:
4.
If the tube loading mechanism is open,
close the regulating clamp
on the administration set
and remove the set. Press the
Done
soft
key. The channel is now shown as
Out Of Service
.
The
pump cannot be used to deliver infusions, but the
Volume
History
key can be used to retrieve history information. Continue to Step 6.
5.
If the channel fails with the tube loading mechanism in the “closed”
position:
Close the regulating clamp on the administration set.
Press the
Open
key. If the mechanism opens, remove the set.
If the mechanism does not open (or if the device has been powered
off due to depleted batteries), use the Manual Tube Release to
remove the set (see “Using the Manual Tube Release,” 4-71).
Figure 8-2 Channel Failure Status
Line Display
Note:
Failures
occurring on
battery power
If the pump is running on battery power and a failure
occurs, the failure message may not be displayed on
the pump module display. The failure will be displayed
on the Main Display, the ALARM LED will light, and the
audible tone will occur.
Note:
Using Manual
Tube Release
Use Manual Tube Release only when the Tube Load-
ing Mechanism is NOT functioning, or if a channel fail-
ure occurs.
Note:
803:07 Failure
Code
If Failure Code 803:07 occurs, ensure that the slide
clamp has been removed from the pump. Do not cut
the tubing to remove the administration set from the
channel. If the tubing is cut, remove the slide clamp
immediately.
! WARNING !
Figure 8-3 Channel Failure
Pop-up Message
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...