Standby Mode
4-52
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
The screen changes to the Main Display and shows
Standby
where
the program status information is normally displayed (Figure 4-75).
If a label was selected while programming the standby infusion, the
label is also shown.
Standby Activation (Triple Channel Pumps)
Note:
Channels B and C are on Standby when the device is first
powered on.
1.
Ensure the pump channel is stopped. Press the
Channel Select
key
for the channel to be put on Standby.
2.
(Optional) To preprogram the channel for future use, program the
infusion (but do not press the
START
key).
3.
Press the
Channel Select
key again. The Standby pop-up is
4.
Press the
×
key next to the
YES
shown on the pop-up to place the
channel into Standby mode.
Figure 4-75 Standby Main Display
Figure 4-76 Standby Pop-up
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...