Programming the Pump
63
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
15. At the Select Infusion screen:
Press the
SECONDARY
soft key to return to the secondary setup. See Figure 54. This
allows reprogramming of the last secondary infusion or programming with a new drug.
See Figure 55.
Press the
review PRIMARY
soft key to return to the primary infusion. See Figure 54.
16. After reviewing the primary program, remove clamp on primary IV line, if applicable, and press
RUN
.
NOTE:
When the secondary infusion is complete, close the clamp on the secondary IV line
above the upper Y injection site or remove the IV bag and tubing to prevent any
remaining drug in the secondary bag from being delivered at the primary delivery rate.
Upon completion of the secondary infusion and once the transition is made to the primary
infusion, a Primary icon (one IV bag) replaces the Secondary icon (two IV bags) on the
Infusion Running screen.
17. In the event the secondary IV bag is still attached, confirm flow from primary drip chamber
(
yes/no
).
If the
yes
soft key is pressed, Primary infuses as programmed.
If the
no
soft key is pressed, close clamp on secondary IV line. Press
OK
to confirm
secondary IV line is clamped.
Clearing the Secondary Infusion to Return to the Primary
Infusion
To clear the secondary infusion to return to the primary infusion:
1. Press
STOP
.
2. Press the
clear program
soft key.
Figure 53. Secondary Complete
Alarm Screen.
Figure 54. Select Infusion
Screen.
Figure 55. Secondary Infusion
Drug Selection Screen.
Содержание 35700BAX2
Страница 152: ...146 Appendix B Flow Rate Accuracy Manual 41018v0800 SIGMA Spectrum Infusion System Revision B Operator s Manual ...
Страница 154: ...148 Appendix B Flow Rate Accuracy Manual 41018v0800 SIGMA Spectrum Infusion System Revision B Operator s Manual ...
Страница 176: ...170 Glossary Manual 41018v0800 SIGMA Spectrum Infusion System Revision B Operator s Manual ...