Product information
PD-TaD 80
20 / 74
822-188-3
No.
Element
Function
7
Anti-corona hoods of the
anti-corona protection
device
Used to protect against corona discharges
8
HV connection cables,
length 0.7 m and 1.2 m
Are used to connect the PD-TaD 80 to the test object
9
Calibrator CAL1B/CAL1E Is used to calibrate the PD test circuit
10
Power Box
(for standalone systems
only)
Is used for:
Power supply for the system components
Data transmission between the PD-TaD 80 and the laptop
11
Connection cables
Are used to connect the system components and the power supply, and for
data transmission between the system components
Further information: Chapter
Connection cables
(on page 22)
3.6.1
Calibrator
Before you can perform a partial discharge measurement, the test circuit must be calibrated
to the following parameters:
Charge (in nC)
Cable length or speed of the PD pulses.
The delivered calibrator is connected to the test object for the calibration.
For information on calibrating the PD measurement system, refer to
the user manual for the BAUR software.