FAQ & General Troubleshoo ng
BATTERY STARTS THEN STOPS WORKING
If the ba ery stops func oning you can reset its internal circuitry by using
the instruc ons below. If the ba ery con nues to shut down, the device
you are a emp ng to power may not be compa ble.
The device may have higher amperage requirements (over 8A) than
the ba ery can provide.
The device may require a constant voltage output to operate
properly. The Freedom™ ba ery has a voltage range of 9V-12.6V DC
and does not maintain a constant voltage level.
The internal circuitry of the device may be set to automa cally adjust
to accommodate for mul -voltage capabili es which could make it
compa ble with the ba ery at mes and incompa ble at other mes.
RESETTING THE INTERNAL CIRCUITRY OF THE BATTERY
If the ba ery stops func oning you can reset its internal circuitry by
holding down the charge level indicator bu on for a full 10 seconds. The
digital display screen will fl ash and two dash marks will appear. Release
the bu on and wait un l the screen goes blank for the circuit board of
the ba ery to reset.
PAP DEVICE CONNECTION VIA A 12V DC POWER CORD
While it is possible to connect a PAP device to the Freedom™ ba ery
using the 12V DC power cord supplied by the device manufacturer and
the DC input cord included with the ba ery kit, the device may only run
for a short period of me. This can occur because many DC power cords
require a constant 12V power fl ow in order to work properly and the
Freedom™ ba ery func ons on a voltage range of 9V-12.6V DC.
TECHNICAL SUPPORT
If you have any ques ons regarding the use of the Freedom™ ba ery
please contact us at 1-877-445-5228 or info@ba erypowersolu ons.net.
Do not a empt to service the Freedom™ CPAP ba ery or any connec on
or charging components yourself.