2-2
920512-768
Operator Manual
Installation Verification
2.2.2 Permissible
Environmental Conditions
This washer is designed to give optimal results under the following
conditions:
Indoor use only;
Altitude of operation up to: 6,267 ft (2,000 m);
Maximum relative humidity is 80% for temperatures up to: 88ºF
(31ºC) decreasing linearly to 50% relative humidity at 104ºF
(40ºC);
Pollution degree 2.
2.2.3 Noise Level
Equivalent Sound Pressure Level at work station, measured 3 ft (1 m)
away from equipment and at 5 ft (1.6 m) from ground:
76.6 dB (A)
.
2.2.4 Seismic Anchorage
System
A Seismic Anchorage System is available for high risk seismic zones
(see
Seismic Anchorage Report 920-512-776
for washer and
920-
512-769
for mechanical core).
2.3 Installation
Checklist
After installing the washer according to the
Uncrating/Installation
Instructions
(P920512-767), complete the following checklist to help
ensure complete and correct installation, or contact STERIS to
schedule a technician to test your installation and demonstrate
proper equipment operation.
❑
Shutoff valves (not provided by STERIS), for maintenance pur-
poses and capable of being locked in OFF position only,
installed on steam, air and water lines and in compliance with
local occupational health and safety regulations, as well as elec-
tric and plumbing codes.
❑
Disconnect switches (not provided by STERIS), for maintenance
purposes and capable of being locked in OFF position only,
installed in electrical supply lines near the unit and in compli-
ance with local occupational health and safety regulations, as
well as electric and plumbing codes. Disconnect switches must
bear the identification of the equipment to which they are con-
nected.
NOTE: If unit is installed next to other equipment, shutoff valves
and disconnect switches should be located so service can be
shut off to one piece of equipment at a time.
❑
Washer is positioned as shown on equipment drawing, with
required clearance space and in relation to building supply
lines.
❑
Washer is level. Remove cabinet panel and adjust leveling legs if
necessary.