29 

 

Redox Calibration 

 

1)  Rinse the electrode with distilled water, then immerse it in the calibration solution (e.g. 220 mV) 

 

2)  Press the CAL key to enter the menu mode and use the 

 / 

 

keys to select the option “IN2 CALIBRATION”  

3)  Press CAL again to confirm 
4)  Press 

  to  select  the  OFFSET  calibration  and  confirm  by 

pressing CAL 

5)  The system automatically recognizes and displays the solution 

value (220 mV)  

6)  If necessary, use the 

 / 

 keys to adjust the calibration value 

7)  Press CAL to confirm the calibration, or ESC to quit the procedure and keep the previous calibration 

 

Notes 

 

If the system does not automatically recognize the buffers or the “Calibration Impossible” error occurs, it can be due to:
 

a) calibration solution contaminated or expired 

 

 

b) electrode faulty or dead 

 

 

c) connection cable or connector damaged 

 

 

During normal operation, it is possible to view the offset value by pressing 

, to check the electrode status. The ideal 

offset value  is close  to  zero. When this value is close to the max / min limits (-100mV ... +100mV), the electrode is 
contaminated or dead. 

Manual Mode 

At any time you can force a manual working mode, useful for a temporary use of the system.  
If a password has been set (see parameter P16), the system will require it to enable the access to this feature. 

1)  Press the CAL key to enter the menu mode and use the 

 / 

 

keys to select the option “MANUAL FUNC.1” (or “MANUAL 
FUNC.2”) 

2)  Press CAL again to confirm 
3)  Similarly proceed for pump 2 
4)  Press ESC at any time to exit the manual mode 

 

Warning!

 In manual mode only one pump at a time can be activated. 

 

 

CONTROL EXAMPLES 

Typical application in swimming pool control: acidification when pH exceeds the pH value of 7.30. 
Refer to the “List of configuration parameters” and set: 

-

  P01 

 ON/OFF working mode for pump 1 

-

  P02 

 set-point 7.30 pH 

-

  P04 

 dosage direction “ACIDIFICATION” 

-

  P05 

 if the quantity of the product to be injected is not known, it is recommended to start with a low working frequency 

(e.g.  20  pulses/minute),  and  if  the  acidification  process  would  last  too  long,  increase  the  frequency  gradually  until  a 
conditioning time of approx. 30-45 minutes 

-

 

P11 

 alarm pump 1 = 60 minutes

 

 
Typical  application  in  swimming  pool  control:  chlorination  when  redox  potential  falls  below  680  mV.  Refer  to  the  “List  of 
configuration parameters” and set: 

-

  P06 

 PROPORTIONAL working mode (recommended for an easier stabilisation) 

-

  P07 

 set-point 680 mV 

-

  P09 

 dosage direction “CHLORINATION” 

-

  P10 

 if the quantity of the product to be injected is not known, it is recommended to start with a low working frequency 

(e.g.  20  pulses/minute),  and  if  the  chlorination  process  would  last  too  long,  increase  the  frequency  gradually  until  a 
conditioning time of approx. 30-45 minutes  

-

  P12 

 alarm pump 2 = 60 minutes 

 
Common settings for the two examples described above: 

-

 

P13 

 start-up delay of 15 minutes (average time required for the polarization of redox electrodes) 

-

 

P16 

 protection PASSWORD to prevent unauthorized access 

 

M  A  N  U  A  L   

F  U  N  C  .   

P  1   

C  A  L  >  Y  E  S   

 

 

E  S  C  >  N  O 

 

M  A  N  U  A  L     P  U  M P     1         
O  F  F   

 

 

0  i  /  m  i  n   

 

 

 

 

I  N  2     C  A  L  I  B  R  A  T  .   

R  X 

C  A  L  >  Y  E  S   

 

 

E  S  C  >  N  O 

 

I  N  2     C  A  L  I  B  R  A  T  .   

R  X 

   

O  F  F  S  E  T   

 

 

 

 

 

 

 

Содержание Tebas-Economic EFka300 pH/Chlor

Страница 1: ...1 Tebas Economic EFka300 pH Redox EN INSTALLATION AND MAINTENANCE INSTRUCTIONS...

Страница 2: ...has to be handled by qualified personnel only Always use original spare parts for maintenance Failure to follow instructions can cause damage to the equipment and in extreme cases to people The manuf...

Страница 3: ...2 PRIMING 23 ELECTRICAL CONNECTIONS 23 Level Control 23 Flow Control 23 Internal Connections for technical personnel only 23 START UP 24 VISUALIZATIONS 24 OPERATION OF THE ELECTRONIC UNIT 25 Standard...

Страница 4: ...in the Technical specifications section The unit must be installed on a vertical wall and securely clamped Choose a location that allows easy calibration operations use and maintenance Place the tanks...

Страница 5: ...polyester front panel Installation wall installation with supplied screws and stoppers Protection Rate IP65 Dimensions 380 x 350 x 175 mm Fixing Template 356 x 250 mm Weight approx 5 kg Dimensions FUN...

Страница 6: ...o the Configuration Calibration and Manual Mode menus confirms parameter modifications Key Scrolls the available menus options in calibration and configuration modes decreases the displayed value in m...

Страница 7: ...es The foot filter must be located at a minimum distance of 5 cm from the tank bottom If a dense product is dosed it is recommended to remove the small inside filter from the foot valve in order to fa...

Страница 8: ...ors while inputs for level sensors one for each pump and flow sensor are available on special connectors that allow an extremely fast and easy connection even for unskilled personnel Note the pH and r...

Страница 9: ...ngly with the configured working mode VISUALIZATIONS During normal operations the display shows the two measured values on the top line while the bottom row displays the status of the two pumps In mod...

Страница 10: ...d related explanations see the List of Configuration Parameters section Standard Configuration The standard configuration mode allows the customer to set a series of parameters related to the normal f...

Страница 11: ...ot protected by password highlighted in bold in the table above PARAMETER 01 FUNCTIONING TYPE OF THE PUMP 1 This parameter allows to set the functioning type of pump 1 typically used for the pH contro...

Страница 12: ...ALARM OF PUMP 2 See parameter P11 but referred to pump 2 PARAMETER 13 START UP DELAY This parameter allows to set a start up delay which is a time of waiting before starting measurements to allow the...

Страница 13: ...pressing CAL 5 The system automatically recognizes and displays the buffer value 7 01 pH 6 If necessary use the keys to adjust the calibration value 7 Press CAL to confirm the calibration or ESC to qu...

Страница 14: ...s ESC at any time to exit the manual mode Warning In manual mode only one pump at a time can be activated CONTROL EXAMPLES Typical application in swimming pool control acidification when pH exceeds th...

Страница 15: ...stem is unplugged Dosing Pumps For a perfect seal after about 800 hours of work it is recommended to tighten the bolts of the pump by applying a torque of 4 Nm Weekly operations Check the level of the...

Страница 16: ...8019 2231 SDE Transparent methacrylate down flow probe holder with standard connections for 8x12 tubes and sampling needle 8061 0252 Flow control kit for SDE electrode holder 9700 9203 EFK1 SGV Compl...

Отзывы: