19 

 

PART 3. THE USB INTERFACE 

The USB interface is used for docking calibration, for data transfer to a computer and for 

setting operating parameters. The cable required for the connection is a six-foot-long, 4-pin, mini 
BM plug to standard AM plug and is available from Bascom-Turner (Part No. UC-001). 

Docking calibration is described in a separate manual and by prompts on the dock. Data 

transfer is described below. 

DataLink4Access™. 

DataLink4Access™ is a software package available from Bascom-

Turner  (Part No. DL-001) for downloading and  archiving data,  viewing and optionally printing 
reports of  data  logged  by  Gas-Explorers,  and  for  maintaining  the  detectors  by  uploading  setup 
parameters.  DataLink4Access™  can  be  installed  on  PC’s  with  Windows  10  or  Windows  7 
operating systems and equipped with a USB port.  See the DataLink4Access™ User Manual for a 
detailed description of the software’s operation. 

 

Figure 3. 

Main Menu, the initial DataLink4Access™ screen

 

1.

 

DOWNLOADS 

The  main  menu  is  structured  into  three  blocks:  Imports,  Reports  and  Maintenance.  The 

Imports block allows selection of three different downloads from the detector to the PC: Readings 
and  GPS  data,  Calibration  data,  and  Unit  data.  Five  report  menus  can  be  selected:  Readings, 
Readings with GPS by Unit, Readings with GPS by Address, Calibration and Units reports. Three 
Maintenance menus allow assigning a user name, resetting date and time and retiring or activating 
detectors.  The Admin button controls functionality of the datalink program itself.  The data storage 
tools enables centralized data back-up and restoration using a network drive. 

Readings.  

Readings are time and date-stamped gas concentration measurements  (e.g. 

%LEL,  %GAS,  %Oxygen,  and  ppm  CO)  recorded  by  the  detector  along  with  identifying 
information such as the detector’s serial number, user’s name, the operational mode used to acquire 
the readings, and the state of the pump when the readings were taken. When the memory is filled, 
the oldest data is automatically overwritten by the new data. 

 

 

Содержание GAS-EXPLORER II EGI-201

Страница 1: ...BASCOM TURNER INSTRUMENTS GAS EXPLORER II DETECTORS OPERATION MANUAL NATURAL GAS CARBON MONOXIDE OXYGEN HYDROGEN SULFIDE Part Number OM 0220...

Страница 2: ...ndise authorization RMA number must be obtained prior to returning a detector to Bascom Turner A detector returned to Bascom Turner with probe and water block filter shipping prepaid will be repaired...

Страница 3: ...from Other Gases Liquids or RFI 9 6 Change of Batteries 10 7 Probes 11 8 Routine Maintenance 12 Appendix 1 Main Sequences for Operation Modes 14 PART 2 SETUP AND CALIBRATION 1 Setup 15 2 Further Setup...

Страница 4: ...sor can be calibrated with System Gas Measurements Over the Full Range of Gas The instruments detect methane over the full range of methane concentrations up to 100 GAS Peak and current readings are d...

Страница 5: ...onds Response Time CH4 0 6 seconds Track Gas 2 0 seconds Monitor H2S CO 45 seconds O2 60 seconds Operating Temp 20o C to 40o C 0o F to 105o F Storage Temperature 40o C to 60o C 40o F to 140o F Data St...

Страница 6: ...s es whose concentration exceeds preset limits Visual alarms are displayed in between readings For example the display will flash CO approximately every two seconds in between normal Monitor mode disp...

Страница 7: ...2...

Страница 8: ...ut a filter on the probe Operation without this filter will eventually degrade the sensors and the pump It also voids the limited warranty It is essential that the filter be replaced monthly or sooner...

Страница 9: ...probe Low gas concentrations at one spot do not necessarily mean that the gas concentration is low throughout a wider area Safety These instruments measure the concentration of natural gas in air and...

Страница 10: ...Probe is displayed followed by Remove Probe and Block Intake and finally Repair Pump followed by an automatic turning off of power The next display suggests a Calibrate by a date thirty days from the...

Страница 11: ...concentration of gas increases according to the chart below 2 Beeps 20 ppm 100 ppm 3 Beeps 100 ppm 250 ppm 4 Beeps 250 ppm 500 ppm 5 Beeps 500 ppm 1 000 ppm 6 Beeps 1 000 ppm 5 000 ppm Since methane...

Страница 12: ...ation of natural gas in nitrogen with a resolution of 0 1 by volume from 0 0 to 100 0 as sensed by the thermal conductivity sensor Purging air filled lines into service requires lowering the oxygen co...

Страница 13: ...rt No FP 012 Air Free CO Flue Measurements The EGA 411 412 and 611 612 detectors can display measurements of CO on an air free basis if this feature is selected see Part 2 When this selection is made...

Страница 14: ...connections block the probe inlet to observe Flow Blocked on the display If Flow Blocked fails to appear there may be a leak Automatic Zero The zero is adjusted automatically when Zero is chosen from...

Страница 15: ...er in thermal conductivity from air examples are hydrogen helium carbon dioxide other hydrocarbons Gases or vapors which may interfere with carbon monoxide detection include substances which can be el...

Страница 16: ...ap and give it one quarter turn battery cap bar vertical Note that the batteries are inserted with the positive terminal first If the instrument does not operate after battery replacement there is a h...

Страница 17: ...ide holes Bar Hole Probe Part No BP 136 36 long fiberglass one hole at end Bar Hole Probe Part No BP 236 36 long fiberglass side holes Bar Hole Probe Part No BP 536 36 long steel side holes The bar ho...

Страница 18: ...ers are not a substitute for dust and water block filters as they will not block water or remove very fine dust and dirt If this filter gets blocked it may be cleaned with compressed air or replaced W...

Страница 19: ...erial to bottom of threads and place the plastic retainer disk flat on top of the filter material Screw in top until finger tight and check seal by carrying out a block test with the detector Heavy Hy...

Страница 20: ...MAIN SEQUENCES FOR OPERATION MODES Off Yes Monitor Track Barhole Purge No Zero Off Mode Monitor CO O2 Zero Off Mode Track PPM peak Off Mode Bar Hole BH Go Off Mode Purge Gas CO O2 Zero Off Mode Monito...

Страница 21: ...s which allow limits to be saved or edited are LEL Alarm Limit 1 to 20 Factory Setting 20 GAS Alarm Limit 0 1 to 1 Factory Setting 1 PPM CO Alarm Limit 1 to 200ppm Factory Setting 35ppm PPM H2S Alarm...

Страница 22: ...d probe are attached The Pump Noise test follows when the Exit key is pressed Pump Noise is meaningful only under load i e when hose probe and filters are connected to the detector s intake nozzle The...

Страница 23: ...l the user can select a One Gas calibration which only uses the 2 5 Methane with 100 ppm CO or the Two Gas calibration which uses both the 2 5 Methane with 100 ppm CO and Pure Methane or System Gas Th...

Страница 24: ...3 Main Menu the initial DataLink4Access screen 1 DOWNLOADS The main menu is structured into three blocks Imports Reports and Maintenance The Imports block allows selection of three different downloads...

Страница 25: ...utes operated since last calibration Number of times the unit was powered on since last calibration Methane sensor cleaned during calibration Y N Unit Data The Unit data is stored in the detector as a...

Страница 26: ...can then be downloaded The Readings and Alarm Events tables can be sorted by unit serial number date time and the operational mode the unit was in when the readings were acquired in order to select th...

Страница 27: ...e choices will filter the data accordingly This new form shows the filtered data and allows plotting the selected data directly in Google Earth A detailed view of the readings and the ability to expor...

Страница 28: ...port of the date and time of each calibration with before and after calibration readings of each sensor Since the pump must work in order to calibrate a detector this is checked first and is indicated...

Страница 29: ...Unit Settings Report gives a listing for the Model number Calibration due time frame Unit and Alarm settings 3 MAINTENANCE Update User Name This will store the user name to the instrument Update Date...

Страница 30: ...back of the case see Step 2 above removing the manifold from the bottom of the case and unscrewing the four screws which secure the pump gasket pump plate diaphragm and piston to the manifold Usually...

Страница 31: ...f case and main circuit board removed CAUTION After replacing a methane or a CO sensor perform the New Sensor procedure described under Part 2 Setup and Calibration and then completely calibrate the i...

Страница 32: ...e filter DS 001 Replacement Dust Filters 5 pkg DF 105 Flue Gas Filter 5 pkg FF 005 Extended Duty Flue Gas Filter 5 pkg FF 105 Heavy Hydrocarbon Filter 5 pkg HF 005 Inlet Filter Quick Connect Air Intak...

Страница 33: ...27 Bascom Turner Instruments 111 Downey Street Norwood MA USA 02062 781 769 9660 800 225 3298 Fax 781 769 2099 bascomturner com...