9. MS3 Meter Visual Check

The MS3 meter should now be operating normally, and the LED at the computer end of the body 
of the meter should be flashing green. 

If not, then follow the guide below:

LED status

Problem

Action

No LED 
indication

No power supply; 5V supply not 
connected to computer side of 
the MS3.

Check the condition of the USB 
cable, and that it is properly 
connected both ends.

If using the RS-232 special cable, 
ensure it is connected to the 5V 
mains adaptor.

Steady orange

The MS3 is communicating with 
the computer, but there is no 
sensor detected.

Ensure that a sensor is connected. 
Check the condition of the cable, 
and that the connectors are tight.

Ensure the drivers have been 
installed on the  computer and are 
working (go to 

Control Panel

 → 

Device Manager

).

Steady red

The MS3 is not communicating 
with the computer.

Check that the USB cable is 
properly connected and inspect the 
cable visually for breaks.

Ensure the drivers have been 
installed on the  computer and are 
working (go to 

Control Panel

 → 

Device Manager

).

When using the RS-232, Bartsoft 
will not operate if the MS3 protocol 
is set to 

A

. Ensure that the MS3 

has been set to either 

B

 or 

C

 

communication mode.

If using the RS-232 special cable, 
ensure it is connected to the 
5V mains adaptor, and that it is 
switched on.

Содержание MS3

Страница 1: ...Operation Manual for MS3 Magnetic Susceptibility Meter...

Страница 2: ...ceptibility Sensor 7 6 4 Connecting to a MS2WFP Power Supply Unit 7 7 Device Drivers 7 7 1 Installed Drivers 7 7 2 Driver Update and Troubleshooting 7 8 Using a MS3 with Bartsoft 8 8 1 Using a USB Cab...

Страница 3: ...is ignored Caution Indicates a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Identifies items that must be disposed of safely to prevent...

Страница 4: ...nvironment Windows or Windows Mobile Software is supplied to operate the MS3 The MS3 meter complies with EMC standards and therefore has a high immunity to magnetic fields in close proximity For the t...

Страница 5: ...ted 5V DC supply socket which is in turn connected to the special 5V mains adaptor all of which are supplied Figure 2 Figure 3 See Connecting to a Computer Serial Port 1 2 3 4 Key to Figure 2 1 Serial...

Страница 6: ...an be downloaded from the Bartington Instruments website The main software designed for the MS3 is Bartsoft If using the MS3 with the Temperature Susceptility system then GeoLabsoft GeoLab should be d...

Страница 7: ...a MS2 Magnetic Susceptibility Sensor Connect the sensor side of the MS3 meter to the MS2 sensor with the TNC TNC cable supplied 6 4 Connecting to a MS2WFP Power Supply Unit Connect the computer side...

Страница 8: ...now working properly with the MS3 meter connected If this still isn t the case Uninstall Bartsoft by going to Control Panel Programs and Features and selecting the program for un installation When pr...

Страница 9: ...ch only recognise serial communication such as AMS Winbar or GeoLabsoft for the MS2 Susceptibility Temperature systems The special RS 232 cable supplied is used to connect the MS3 meter to the serial...

Страница 10: ...rt displayed and set the port settings to match that of the protocol setting B selected on the MS3 1200 bps 8 data bits no parity and 2 stop bits Click on Advanced and drag both Transmit and Receive b...

Страница 11: ...upply is 5V rather than 12V Using a higher voltage may result in a protective fuse being blown in the MS3 which will need replacement by Bartington Instruments On the computer go to Control Panel Devi...

Страница 12: ...computer but there is no sensor detected Ensure that a sensor is connected Check the condition of the cable and that the connectors are tight Ensure the drivers have been installed on the computer and...

Страница 13: ...n 44 0 1993 706565 Caution Fault finding by customers may invalidate the warranty The meter is not user serviceable The most likely causes of failure and their solutions are detailed in MS3 Meter Visu...

Страница 14: ...BARTINGTON INSTRUMENTS Page 14 of 15 OM3227 2 Notes...

Страница 15: ...Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thorney Leys Business Park Witney Oxford OX28 4GE England www bartington com T...

Отзывы: