OM0065 ISSUE 5  PAGE 5 of 21  

 
5  CABLES 

      
Cables can be supplied to connect any of the 

Mag

-03 series of  sensors to the 

Mag

-03PSU.  

Details are given in the operation manual for the sensor.  The cable carries the supply voltage to 
the magnetic field sensor;  0 V connection is via the screen.  Analog signals are returned via three 
conductors and the analog return is via a common conductor.   
      
The maximum cable length of 600 metres is derived from the following considerations: 
      
(a)   

Above this length the positive supply current of 20 mA will produce a 1.5 V voltage 

 

drop which may restrict the analog output swing especially where a low voltage supply is 

 

used.      

 
(b)   

Above this length the maximum capability for capacitance loading of the magnetic field       

 

sensor output will be exceeded. 

      
Cable type 
      
Cables other than those provided by Bartington Instruments should generally conform to the 
specifications given and a screened type is essential if low noise performance is to be achieved.  
An example of an alternative cable is Belden Type 9536 for use in protected conditions, e.g. ducts 
etc. 
      

 
6  ANALOG VOLTAGE MEASUREMENTS 

      
The low resistance overall cable screen is used as a power 0 V return and, in addition to the three 
X, Y and Z analog outputs, a separate  signal common line is provided.   The provision of a 
separate signal common line ensures that the voltage developed in the power return, due to the 
power supply currents, does not appear in the analog measurement.  Clearly this is only possible 
where no electrical connection is made between the power supply and the analog output common 
connection at the measurement and power supply end of the cable.  Connection of the output 
signals to an isolated measurement device such as a battery powered voltmeter will  introduce no 
errors.  Connection of the analog outputs to, for example, a data logger may provide a source of 
error if the analog input LOW is connected to the electrical mains supply.  Most data loggers have 
differential inputs and this eliminates completely any ground-loop problems.  It is important, 
however, when using a differential input measurement instrument to constrain the common line 
voltage to some small value.  This is usually accomplished by providing a single ground 
connection at the power supply. 
      
To obtain the full precision of the magnetic field sensor the analog outputs should be connected 
only to analog inputs which have an impedance greater than 10 k

 
 

7  

Mag

-03PSU WITH 

Mag

-03 SERIES SENSORS 

    
Use of the 

Mag

-03PSU power supply eliminates any problems which may be encountered in 

providing a suitable isolated power source.  The unit also provides additional filtering of the 
analog signal to improve the signal to noise ratio.  The high pass filter section can be selected to 
provide either a dc or ac response. 
      

Содержание MAG-03PSU

Страница 1: ...3PSU POWER SUPPLY UNIT OPERATION MAINTENANCE MANUAL Distributed by GMW Associates 955 Industrial Road San Carlos CA 94070 USA Tel 650 802 8292 Fax 650 802 8298 Email sales gmw com website http www gmw...

Страница 2: ...ALARM 10 ANALOG OUTPUT 11 TAKING MEASUREMENTS 11 1 CONNECTING UP 11 2 FILTER SECTION 11 3 ADDITIONAL FILTERS 11 4 OVERLOAD 11 5 AC MEASUREMENT PRECAUTIONS 11 6 ELECTROMAGNETIC COMPATIBILITY 12 DETAIL...

Страница 3: ...CHEMATIC FIGURE 2 Mag 03PSU ASSEMBLY FIGURE 3 PC36 SCHEMATIC Mag 03PSU FIGURE 4 PC36 ASSEMBLY AND FILTER SELECTION FIGURE 5 Mag 03PSU WIRING DETAILS PARTS LISTS PL0065 Mag 03PSU ASSEMBLY PL0063 PC36 A...

Страница 4: ...which is used for the 0 V power supply return common line See Figure 1 3 FILTERS A low pass and a high pass filter are included in each of the channels for the signals from the X Y and Z axes of the s...

Страница 5: ...a separate signal common line ensures that the voltage developed in the power return due to the power supply currents does not appear in the analog measurement Clearly this is only possible where no...

Страница 6: ...dible alarm will be heard to sound continuously in which case the battery should be recharged immediately The low battery alarm will sound for a short time when the power supply is switched on to conf...

Страница 7: ...hould be limited using an external filter with a sharp cut off to the frequency band of interest 11 4 OVERLOAD Exposure of the magnetic field sensor to a flux between 120 and 200 or 120 and 400 of the...

Страница 8: ...cant dc offset error due to circulating current For this reason the analog signal return is via a separate conductor appearing on Pin 11 of PC36 The analog return is shared by X Y and Z channels To pr...

Страница 9: ...cable disconnect it and see if the condition persists iii If the condition still exists then the voltage converting and filtering circuitry should be suspected iv If the cable is suspected use a conti...

Страница 10: ...following test Connect the 6 8ohm resistor and voltmeter simultaneously across the battery terminals A battery in reasonably good condition should sustain a voltage of 5 8 V minimum for five seconds F...

Страница 11: ...epending on condition and state of charge of batteries Alternative YUASA NP 1 2 6 Battery alarm sounds when battery voltage less than 5 V and at switch on Internal charger temperature compensated cons...

Страница 12: ...OM0065 ISSUE 5 PAGE 12 of 21 DR1011 1 Mag 03 Mag 03PSU SIMPLIFIED SCHEMATIC FIGURE 1...

Страница 13: ...OM0065 ISSUE 5 PAGE 13 of 21 DR0065 7 Mag 03PSU ASSEMBLY FIGURE 2...

Страница 14: ...OM0065 ISSUE 5 PAGE 14 of 21 DR1013 2 PC36 SCHEMATIC FIGURE 3...

Страница 15: ...OM0065 ISSUE 5 PAGE 15 of 21 DR1014 1 PC36 ASSEMBLY FILTER SELECTION FIGURE 4...

Страница 16: ...OM0065 ISSUE 5 PAGE 16 of 21 DR1015 2 Mag 03PSU WIRING DETAILS FIGURE 5...

Страница 17: ...RACTICAL CM3 150 G RFI 1 POWER 9 PILLAR 3 X 26mm HAWNT NPB 26 B M3 3 10 DOUBLE SIDED TAPE 1 100mm LONG 11 M3x10mm CSK POZI SCREW 3 STEEL BRIGHT ZINC PLATED CLEAR PASSIVATED 12 M3 PLAIN STEEL WASHER 9...

Страница 18: ...ONG BLACK 20 No 6 SELF TAPPING SCREW 1 6 5mm LONG STEEL BRIGHT ZINC PLATED CLEAR PASSIVATED 21 LOCTITE 7400 AS REQD 22 STICK ON FEET 4 POLYEURETHANE SELF ADHESIVE 12 7 X 3 6 BLACK 23 FR4 NUT 1 USED TO...

Страница 19: ...5Q 3 12 C12 13 14 CAP 1 0 F POLYESTER MAPLINS WW53H 3 13 C5 7 CAP 22 16 TANTALUM FARE 100 882 2 14 C6 8 CAP 68 16 TANTALUM FARE 100 885 2 15 C3 CAP 100 F 10V TANTALUM FARE 100 875 1 16 L1 2 0 47mH IND...

Страница 20: ...GREEN X 150mm 1 34 CABLE 7 0 2 BLUE X 150mm 1 35 CABLE 7 0 2 BLACK X 150mm 2 36 CABLE 7 0 2 GREEN YELLOW 1 TO COMPLETE ESD SHIELD X 100mm 37 M3 SOLDER TAG 1 38 4 8 mm BLACK HEATSHRINK 1 X 10mm LONG 2...

Страница 21: ...NUT RSCO 560 271 2 8 LED RED RSCO 826 492 1 Is 1 6mA ALGAAs 9 MOUNTING BEZEL RSCO 589 569 1 10 PTFE SLEEVING FARE 583 972 AS ANY CLEAR ALTERNATIVE ALLOWED REQD TO COVER LED WIRES RESISTOR 11 RESISTOR...

Отзывы: