To Replace the Mode Switch
and/or Power Switch
1.
Disconnect mixer from power supply.
2.
While holding the rubber mixing cup securely, loosen
the screw holding the cup down and remove the cup
from the counterweight.
3.
Turn mixer upside down and remove the four screws
securing case. Turn mixer upright and remove case.
4.
Disconnect wires from switch. Identify or mark wires
disconnected to ensure proper placement and con-
nection when reinstalling.
5.
Remove defective switch.
6.
Install new switch with the number “3” terminal on
top. Install the wires as identified or marked in Step
4.
7.
Reinstall case and secure with four screws.
8.
While holding the rubber mixing cup securely,
secure mixing cup to counterweight with the hold
down screw.
9.
Reconnect mixer to power supply.
To Replace Microswitch
1.
Disconnect mixer from power supply.
2.
While holding the rubber mixing cup securely, loosen
the screw holding the cup down and remove the cup
from the counterweight.
3.
Turn mixer upside down and remove the four screws
securing case.
4.
Turn mixer upright and remove case.
5.
Remove four screws and nuts securing motor.
6.
Position motor on its side.
7.
Disconnect the two wires from microswitch and
remove microswitch from motor. Save the spring and
9
Maintenance and Servicing
Warning
To avoid electrical shock, always dis-
connect from power supply before
servicing. Refer servicing to qualified
personnel.
Содержание Maxi Mix Plus M63210-33
Страница 15: ...15 ...