Barnstead International Harvey Hydroclave MC8 Скачать руководство пользователя страница 1

1

Harvey

®

Hydroclave MC8

Hydroclave MC10

Steam Sterilizer

(With Optional Non-Recirculating System)

OWNER AND OPERATOR MANUAL

LT1203X1 • 5/11/04

Содержание Harvey Hydroclave MC8

Страница 1: ...1 Harvey Hydroclave MC8 Hydroclave MC10 Steam Sterilizer With Optional Non Recirculating System OWNER AND OPERATOR MANUAL LT1203X1 5 11 04...

Страница 2: ...m 13 Operation 15 Power On Controls On 16 Filling Reservoir 17 Filling Non Recirculating Tank Option 18 Emptying Collection Bottle 18 Installing Overflow Tube Optional 19 Load Chart 20 Preparing Items...

Страница 3: ...eaning Printer Recess MC10 34 Replacing Door Gasket 35 Every 6 Months 36 Cleaning Non Recirculating Water Cooling Tank 37 Recommended Spare Parts 38 Installation 39 Leveling Sterilizer Cabinet Clearan...

Страница 4: ...e information included in this manual applies to all models unless noted for specific models We urge you to read this manual carefully before operating the unit and retain it for future reference Oper...

Страница 5: ...At the completion of the exposure time a sole noid valve opens allowing the steam and water to vent back into the reservoir The steam reenters the reservoir through tubing below the water level in the...

Страница 6: ...C8 and MC10 steam sterilizer Even when distilled or deionized water is used some contaminants can collect in the chamber rack and trays Clean the reservoir reservoir filter and chamber screen at least...

Страница 7: ...lectrothermal Engineering Ltd 419 Sutton Road Southend On Sea Essex SS2 5PH United Kingdom Copies of the Declaration of Conformity are available upon request This equipment has been tested and found t...

Страница 8: ...ptacle Attention Consult Accompanying Documents Refer to Warning and Caution notes in this manual Type B Equipment Class I Refer to IEC Standard 601 1 OFF Only for a part of equipment ON Only for a pa...

Страница 9: ...r Accessory Optional Drain Tube Open Lock Printer Receptacle MC8 Fuse s Power Cord Discharge Port Non Recirculating Pressure Relief Valve Water Fill Port Chamber Filter Screen Reservoir Drain Valve Do...

Страница 10: ...liquid cycle Drying times are adjustable Cycle runs automatically when START switch is pressed Optional printer records temperature and pressure during cycle Chamber temperature may be displayed in C...

Страница 11: ...11 GENERAL DESCRIPTION Figure 1 Control Panel Harvey Hydroclave MC8 Cycle Phase Indicators Cycle Select Switches...

Страница 12: ...ssories MC8 MC10 Support Flange Tray Rack MC8 Part No 265002 MC10 Part No 2 513025 Large Tray Part No 2 67811 Small Tray Part No 265001 Wrapped Cassette IMS Cassette Not Sold By B T IMS Cassette Rack...

Страница 13: ...ng tank This feature allows the excess water to drain directly into a sink or drain if available eliminating the need to periodically empty the collection bottle Recommended Steam Sterilization Monito...

Страница 14: ...vials Spore Population 106 AY759X8 Process Indicators Chemical Indicator Strips Box of 250 Steam Sterilizers Only AY759X2 Chemitest Indicator Tape Closure system for packs bags or other containers 260...

Страница 15: ...7 Open Door to Dry 8 Remove Load Optional Drying phase if selected will start automati cally without opening the door Opening the door however may improve drying efficiency Operation Operation Warning...

Страница 16: ...temperature exposure time and drying time will be displayed and the Standby Mode will begin During the Standby Mode the heater turns on to warm the cham ber to approximately 38 C The heater then cycle...

Страница 17: ...Caution Use only distilled or low quality deionized water minimum 0 5 megohm cm maximum 1 5 megohm cm Normal tap water contains minerals that will corrode sterilizer parts Warning Burn Hazard Do not...

Страница 18: ...er Cooling Tank Collection Bottle From Sterilizer Figure 8 Non Recirculating Water Accessory Fill With Distilled or Low Quality Deionized Water Until It Overflows Into the Collection Bottle Approx 3 G...

Страница 19: ...latile substances or for any purpose other than its intended design Failure to observe this warning could cause personal injury or damage to equip ment Warning Burn Hazard To prevent the possi bility...

Страница 20: ...s e t t e s s a c S M I e e r h T s b l 5 1 d e p p a r w e l g n i s h c a e g 0 8 6 y t n e w t r o s e h c u o p 0 1 x 6 t h g i E d e g n i h o w t s e h c u o p 5 x 3 h c u o p r e p s t n e m u...

Страница 21: ...terilizer piping This could eventually cause partial restriction of water flow and the need for possible service Unwrapped goods can be placed directly on tray liner in the bottom of the tray Do not u...

Страница 22: ...le Figure 11 Selecting Cycle s t n e m u r t s n I d e p p a r w n U m u m i n i m s e t u n i m 3 s t n e m u r t s n I d e p p a r W m u m i n i m s e t u n i m 5 s k c a P m u m i n i m s e t u n i...

Страница 23: ...cient drying opening the door is strongly rec ommended See Open Door Drying procedure At the end of closed door drying three groups of three beeps will sound and the COMPLETE indicator lights To cance...

Страница 24: ...e minute OPERATION Figure 13 Drying Cycle Operation Latch Handle Down With Door Open Door Slightly Open Approx 1 4 Latching Bars Note Closed door drying will stop when the door is opened When START is...

Страница 25: ...he power interruption If you cannot start a cycle after power is restored 1 Unplug the unit 2 Press and hold the OFF button while plugging the unit back in 3 Release the OFF button after PERFORMING SE...

Страница 26: ...C R E B M A H C k c e h C r o o d n e p o n e h t l o o c o t r e b m a h c r o f t i a W r e t l i f r e b m a h c n a e l c r o r e b m a h c n i s i r b e d r o f f i e c i v r e s l l a C n i a g...

Страница 27: ...e z i l i r e t S f o n o i t e l p m o c r e t f a h g u o n e e l c y c e z i l i r e t s t e w y l e v i s s e c x e d a o L e g r a l o o t e z i s d a o L s l l a w r e b m a h c g n i h c u o t...

Страница 28: ...installed If a printer is added later it should be installed by an authorized service technician Operation The printer provides hard copy documentation of each cycle An ON OFF PAPER FEED switch is pro...

Страница 29: ...move the ribbon cover press down as shown Lift the cover up and out 2 To remove the ribbon press down as shown Grasp the ribbon and lift it from the printer 3 Press the new ribbon down until it snaps...

Страница 30: ...ss 2 Rest the printer on the top cover while cleaning the recess It is not necessary to disconnect the printer cables 3 When installing the printer tuck the excess cable into the recess PRINTER Figure...

Страница 31: ...er at room temperature Wipe all residue from the surfaces with a dampened lint free cloth Operator Maintenance Figure 22 Cleaning Chamber Rub Area To Be Cleaned Rinse with Water and Wipe Clean Figure...

Страница 32: ...n hose and fitting 4 Store the drain tube and fitting OPERATOR MAINTENANCE Figure 23 Draining Reservoir Warning Water May Be Hot Drain Fitting Click In To Drain Release Tab Push To Release Drain Port...

Страница 33: ...hod does not effectively clean the filter replace it Cleaning Chamber Filter Screen 1 Pull the filter screen assembly out of the water fill port inside the chamber 2 Clean any deposits from the screen...

Страница 34: ...e Use a mild disinfectant to clean the recess OPERATOR MAINTENANCE Figure 27 Emptying Non Recirculating Collection Bottle Collection Bottle Empty Contents Down The Drain Note The fill port is the bott...

Страница 35: ...ter stud of the door plate 4 Insert the retainer inside the gasket outer lip of retainer against gasket as shown 5 Slide the gasket and retainer assembly onto the center stud of the door plate and se...

Страница 36: ...malfunctions the valve must be replaced Call for authorized service 4 Cancel cycle after checking valve to prevent overheating OPERATOR MAINTENANCE Figure 29 Pressure Relief Valve Note If traces of s...

Страница 37: ...household bleach 3 4 cup and distilled or deionized water 1 gal Drain solution and rinse with distilled or deionized water before refilling Clean filter in the inlet line fitting before connecting to...

Страница 38: ...u F D T e p y T 2 6 0 3 4 6 2 m m 0 2 x 5 t l o v 0 5 2 p m A 8 l e d o M t l o v 0 3 2 e s u F D e p y T 2 6 7 0 5 6 2 r e t l i F r i o v r e s e R n e e r c S 1 1 8 0 5 6 2 n i a r D r e b m a h C...

Страница 39: ...rate Circuit Installation Warning Explosion Hazard Do not locate this steril izer in areas where explosive anesthetics are used or stored An explosion could occur causing possible equipment damage and...

Страница 40: ...ing discharge port Step 3 Automatic Optional Overflow Push the overflow tube into the overflow connector at the top of the cooling tank until it clicks Tighten nut finger tight plus 1 4 turn with wren...

Страница 41: ...to set the hour ahead or the down arrow to set it back When the clock is set to the correct hour press PROGRAM SET to store the hour in memory b The minute will flash Use the arrows to set the minute...

Страница 42: ...to store e The day will flash Use the arrows to set the day Press PRO GRAM SET to store f The year will flash Use the arrows to set the year Press PRO GRAM SET to store 3 The clock calendar is now rea...

Страница 43: ...ease the PROGRAM SET switch The sterilizer is now ready for you to proceed with Step 2 INSTALLATION Note This operation requires the use of two hands You must tell the steriliz er that the non recircu...

Страница 44: ...AM SET Press PROGRAM SET Press PROGRAM SET Press PROGRAM SET INSTALLATION LOG ERXX XXXXX DATE TIME PRESSURE UNITS PSI TEMPERATURE CELSIUS 12 HOUR CLOCK DIAGNOSTICS NO LANGUAGE ENGLISH CONFIGURATION HA...

Страница 45: ...SET 3 When RECIRCULATE is displayed press either the up or down arrow to change to NON RECIRCULATE Press PROGRAM SET Press STOP This will revert back to the main display menu time and date STANDARD UN...

Страница 46: ...quarts Maximum Rated 45 psi 310kPa 45 psi 310 kPa Operating Pressure Environmental Conditions Operation 10 to 40 C 50 to 104 F 10 to 40 C 50 to 104 F 30 to 70 relative humidity 30 to 70 relative humid...

Страница 47: ...or save for future reconversion Appendix Figure 31 Removing access panels Power I O Board Terminal Block Transformer PC Board MC10 Printer Option Warning Unplug power cord before removing any access p...

Страница 48: ...er wire from J14 to J11 Leave other black wire connected 3 On the Printer Power PC Board MC10 with Printer Only Remove jumpers from JMP1 and JMP2 positions and install jumper in JMP3 posi tion Save ot...

Страница 49: ...49 49 APPENDIX Figure 32 Power I O PC Board 230 VAC...

Страница 50: ...50 50 Figure 33 MC10 Printer Power Supply Board 230 VAC APPENDIX...

Страница 51: ...rom J11 to J14 Move white heater wire from J12 to J11 Leave the other black wire connected 3 On the Printer Power PC Board MC10 with printer only Remove jumper from JMP3 posi tion and install jumpers...

Страница 52: ...52 52 Figure 34 Power I O Board 115 VAC APPENDIX...

Страница 53: ...53 53 APPENDIX Figure 35 MC10 Printer Power Supply Board 115 VAC...

Страница 54: ...54...

Страница 55: ...55...