INSTRUCTION MANUAL
DPS400
56-101-01180-C
1-4
Page 18
November 5, 2003
(e) Pitot Limits to Zero Airspeed (To set Airspeed to zero see previous section, Setting Pitot Limits)
Note:
The pneumatic regulators in previous versions of the DPS400 were unable
to achieve zero airspeed (0 kts or 0.0 inHg differential pressure) at high
altitudes. The user interface/control of the crossbleed solenoid valve was
modified to ensure against operator damage to the instrument of aircraft-
under-test while zero airspeed is set.
Caution:
Set pumps to ON. Maintain pumps ON when setting zero airspeed or while
testing with the zero airspeed special mode.
(If pumps are turned OFF during
zero airspeed testing, use
F4
to cancel out of special zero airspeed mode.)
If 0 (zero) airspeed is set, the DPS400 shows the following screens:
:
ZERO AIRSPEED LIMIT = 0 Kts
SPECIAL MODE = Zero Airspeed
F1: To Continue
F4: -Cancel-
F1 F2
LEAK
TEST
INC
F3
F4
CONTRAST
DEC
MENU
PUMP
•
Press
F1
to continue. The following screen appears:
CAUTION
Verify at least one of the
following two conditions is true.
F1:
To Continue
F4: -Cancel-
F1 F2
LEAK
TEST
INC
F3
F4
CONTRAST
DEC
MENU
PUMP
•
Press
F1
again. The next screen appears
Condition
(1)
Ps and Pt ports presently
(vented) at Ground
F1:
To
Continue
F4:
-Cancel-
F1 F2
LEAK
TEST
INC
F3
F4
CONTRAST
DEC
MENU
PUMP
•
Press
F1
to continue; (
F4
to cancel and return to the previous screen)