The operation of the solenoid valves can be summarized as follows:
Deviation From
Limit
Exceedance
Programmed
Limit
Solenoid
Valves
Actuated
Altitude
±
1000
ft,
350
m
SV1A,
SV1B
Airspeed
±
50
kts,
100
km/hr SV1A,
SV1B,
SV3A
VSI
±
500
ft/min,
150
m/min
SV1A,
SV1B,
SV2A,
SV2B
Mach
±
0.2
mach
SV1A,
SV1B,
SV3A
Note: It is imperative that the operator monitors the displayed parameters closely to avoid
excessive differential pressures between the test set and the aircraft systems once the
solenoid isolation valves are actuated. Refer to Operations paragraph
PROTECTION CIRCUIT RESET INSTRUCTIONS
of this manual for instructions
on how to equalize static/pitot system pressures and return them to ambient after the
protection circuits have tripped.
5. PNEUMATIC SCHEMATIC DIAGRAM
shows the internal pneumatic connections of the most recent production DPS400
Mod P. For previous configurations, refer to earlier revisions of this instruction manual.
PNEUMATIC DIAGRAM
Figure 6