Sherlock 3CG™ Tip Confirmation System

Step 5: Position Sensor and ECG electrodes

- Remove the adhesive backing from the sensor holder and lace the sensor directly on the patient’ skin with the adhesive side down. Place the sensor as flat as 

possible for best results.

Note:

 The sensor should be positioned the same for left or right side placements.

- Prepare and attach external ECG electrodes per the following steps:  
• 

Ensure electrode locations are oil-free and completely dry.

Caution:

 Electrodes should be applied only to intact, clean skin (e.g., not over open wounds, lesions, infected or inflamed areas).

Attach electrodes to all lead wires.

Remove backing and press electrodes firmly onto skin at the specified locations.

 •  Place 

black

 electrode on patient’s lower right shoulder.

 •  Place 

red

 lead on patient’s lower left side, inferior to the umbilicus and laterally along the mid-axillary line.

Caution: 

Placement of red electrode outside of this region may result in reduced ECG performance.

Warning: 

Place skin electrodes carefully at locations indicated in these Instructions for Use and ensure good skin-electrode contact. Failure to do so may cause 

unstable ECG waveforms and/or ECG waveforms that are not described in these Instructions for Use. In such a case, rely on magnetic navigation and external 

measurement for tip positioning and use chest X-ray or fluoroscopy to confirm catheter tip location, as indicated by the institutional guidelines and clinical 

judgment.

Caution:

 Discontinue electrode use immediately, if skin irritation occurs.

Tips:
- Prior to securing the sensor holder to the patient, it may be necessary to clean the skin and remove excess hair.
- Do not move the sensor after it is secure. Best results will be achieved if the patient remains still and the sensor is not placed on open wounds, over bandages,    

  drapes, gowns or other coverings.

Содержание Sherlock 3CG TCS

Страница 1: ...on date March 2014 Bard Sherlock Sherlock 3CG and Site Rite are trademarks and or registered trademarks of C R Bard Inc All other trademarks are the property of their respective owners 2014 C R Bard I...

Страница 2: ...For use with Site Rite 6 Ultrasound System Instructions for Use...

Страница 3: ...Operating Temperature Parameters Do not dispose with ordinary municipal waste Operating Humidity Parameters Storage Humidity Parameters unpackaged Storage Humidity Parameters packaged Keep Dry Manufa...

Страница 4: ...Tracking Calibration Step 8 Prepare Catheter Sterile Field Step 9 Access the Vein Step 10 Attach Catheter Stylet to Fin Assembly Step 11 Perform Final Magnet Navigation Calibration Step 12 Insert Cat...

Страница 5: ...y positioning the tip of catheters in adult patients The primary endpoints were to assess the performance of the Sapiens technology with respect to 1 the accuracy of the Sapiens technology with regard...

Страница 6: ...m controls for using Sherlock 3CG TCS Mode on the Site Rite 6 Ultrasound System Authorized accessories include Site Rite 6 Ultrasound System Sherlock 3CG TCS Sensor Sherlock 3CG TPS Stylet included wi...

Страница 7: ...MRI devices The high magnetic fields created by an MRI device will attract the equipment with a force sufficient to cause death or serious injury to persons between the equipment and the MRI device T...

Страница 8: ...e roll stand see illustrations below Caution When the sensor is not in use store in the holster roll stand basket or other secure location to avoid damage 2 2 Connecting the Sherlock 3CG TCS Sensor to...

Страница 9: ...7 Sherlock 3CG Tip Confirmation System 3 2 Sherlock 3CG TCS Graphical Interface...

Страница 10: ...8 Instructions For Use...

Страница 11: ...9 Sherlock 3CG Tip Confirmation System...

Страница 12: ...dio On Select to mute the magnet navigation audio Audio Mute Select to turn the magnet navigation audio on Freeze Select to copy the current ECG waveforms from the Main Screen to the Reference Screen...

Страница 13: ...nu Window Patient Information Window Settings Window or quit Demonstration Mode Exit Site Marking Controls Increase Select to cycle through the exit site marking options in a positive progression Decr...

Страница 14: ...he Menu Window Settings Controls P wave Highlighting Select to enable or disable highlighting the P wave on the ECG signals ECG Color Select to enable or disable the intravascular yellow ECG waveforms...

Страница 15: ...ignals are only applicable to the magnetic navigation capability of the device Tone 4 is applicable to ECG in patients where there are no alterations to the cardiac rhythm that change the presentation...

Страница 16: ...Rite 6 Ultrasound System via USB cable Verify that the ECG flatline signal is scrolling Verify the battery charge is sufficient for the procedure Enter Patient Information as needed Step 2 Position P...

Страница 17: ...r to the umbilicus and laterally along the mid axillary line Caution Placement of red electrode outside of this region may result in reduced ECG performance Warning Place skin electrodes carefully at...

Страница 18: ...on As the stylet tip approaches the sensor an icon appears at the edge of the screen indicating the approach of the stylet tip When the stylet is under the sensor the stylet and depth icons will displ...

Страница 19: ...nnect the sensor to the Site Rite 6 Ultrasound System via USB cord Verify the battery charge is sufficient for the procedure Enter patient Information as needed Step 2 Position Patient and Perform Ult...

Страница 20: ...patient to remain still and do not reposition the patient Refer to Bard Access Systems catheter Instructions for Use for catheter insertion Step 10 Insert Catheter Refer to Bard Access Systems cathet...

Страница 21: ...ectrical outlet to run the system on battery power Sensor cannot be detected by the display Call the technical support hotline at 800 443 3385 Interference caused by magnetic field changes or by senso...

Страница 22: ...sor and ask the patient to remain still 5 Ensure catheter tip is at least 12 inches 30 cm away from the sensor 6 Calibrate 7 If the error continues discontinue use and call the technical support hotli...

Страница 23: ...y Patient movement Ask the patient to remain still Ensure the patient is warm and relaxed Extraneous noise Unplug the Site Rite 6 Ultrasound System from AC power Make Sure the Ultrasound probe is not...

Страница 24: ...y using tape If the vein is shallow lift the sensor away from the skin e g by using a folded blanket Too much time has passed since calibration Magnetic fields have changed Pull catheter tip back at l...

Страница 25: ...se Solution Cause Solution Cause Solution Cause Solution Stylet at edge of sensor s depth range 3 to 11 cm 1 Pull catheter back until stylet icon is stable 2 Verify catheter is moving in the correct d...

Страница 26: ...llowing actions void the warranty of the Sherlock 3CG TCS Sensor and may result in injury or equipment damage Opening or servicing the Sherlock 3CG TCS Sensor by anyone other than Bard Access Systems...

Страница 27: ...ures may result in damage to the system Warning Opening or servicing the Sherlock 3CG TCS Sensor by anyone other than Bard Access Systems authorized service personnel will void the warranty and may re...

Страница 28: ...26 Instructions For Use...

Отзывы: