24
Site~Rite* 6
Ultrasound System
5 Settings
5.1
Date, Time and Image Size
To modify date, time and image size settings:
1. Press the Freeze button ( ) to pause the system.
2. Press and hold the Depth button (
cm
) for more than 5 seconds to enter the time/image size mode.
The year field will then be highlighted. Release the Depth button.
3. Press the Depth button (
cm
) to select the desired date or time field to be modified.
4. Press the Gain buttons ( , ) to change the selected date or time values.
5. After setting the date and time, press the Depth button (
cm
) to enter the image size field.
The current image size will be highlighted.
6. Press the Gain buttons ( , ) to select the desired image size.
7. Press the Freeze button ( ) to store the updated settings and resume normal operation.
5.2
Image Parameters
The Site~Rite* 6 Ultrasound System image may be changed from the factory default settings.
Contrast
• Press and hold the Depth Marker button (/:\) for 5 seconds to toggle between the high and low
contrast image.
Smooth/Sharp
• Press the F5 key to toggle between a smooth and sharp image.
Содержание 9770000
Страница 1: ...ACCESS SYSTEMS INSTRUCTIONS FOR USE REF 9770000 ...
Страница 2: ......
Страница 3: ...INSTRUCTIONS FOR USE ...
Страница 8: ... 6 Site Rite 6 Ultrasound System ...
Страница 42: ......
Страница 43: ......