R5906150 /05
XDL
26
There are three cable glands (reference 4
) at the base of the mains compartment for the cabling of
the external interlock devices.
The projector is shipped with a connector of which the external beam enable key and the 2 external interlocks
are bypassed. Before connecting an external interlock the appropriate shunt wires (reference 6
should be removed and replaced by the interlock wiring.
Component location
1
2
3
4
5
6
7
8
9
10
A1
A2
MK
B1
B2
4
1
2
3
5
6
Image 2
–
9
1
Beam stop button.
2
Beam enable key.
3
Socket and connector for external safety interlocks.
4
Three cable glands (M20 D7
–
13).
5
10 pins connector.
6
Shunt wire.
A1
External safety interlock A contact No 1.
A2
External safety interlock A contact No 2.
MK
External beam enable key.
B1
External safety interlock B contact No 1.
B2
External safety interlock B contact No 2.
CAUTION:
The need for installing external safety interlocks is the responsibility of the Safety Officer
and depends on the installed application and local regulations.
WARNING:
The connector with shunt wires is delivered with the projector but not installed. The
Safety Officer of the boot area must decide either if the connector with shunt wires can be installed
or if external safety interlocks must be installed.
The external safety interlocks used must be of the type 'micro load' and must be approved interlocks
for safety applications. In order to achieve the highest reliability level of safety; employ only SIL level
4 (according IEC 62061) dual normal open contact or performance level Ple (according IEC 13849-
1) dual normal open contact devices.
Содержание XDL
Страница 1: ...ENABLING BRIGHT OUTCOMES Installation Manual XDL...
Страница 30: ...R5906150 05 XDL 30 Installation process...
Страница 74: ...R5906150 05 XDL 74 Pulse SFP input use cases...
Страница 79: ...79 R5906150 05 XDL Dimensions A...
Страница 82: ...R5906150 05 XDL 82 Glossary...
Страница 86: ...R5906150 05 XDL 86 Index...
Страница 87: ......
Страница 88: ...www barco com R5906150 05 2021 08 02...