8. Primary pump unit
1
1
Image 8-16
Installing the two screws
3. Tighten the two screws (reference 1 image 8-17) (rear side of the chiller) using the nut spanner 8 mm.
1
1
Image 8-17
Tightening the two screws
4. Plug in the primary pump cable connector.
Image 8-18
Plugging the primary pump Unit cable connector
5. Connect the two open type quick couplers of the refrigeration unit.
Note:
First remove the caps and plugs. Check the cleanliness of the openings.
Warning:
Risk of serious liquid spillage: Ensure that the clips are correctly & fully engaged after assembly. An audible click
can be heard when the connector is correctly inserted.
44
R5906015 ULC-30A 01/02/2018