background image

37

R5913350 /05

Nio Fusion 12MP

To enable/disable image scaling:

1.

Bring up the OSD main menu.

2.

Navigate to the

Configuration > Image Source

menu.

3.

Enter the

Scaling

submenu.

4.

Select

Enabled/Disabled

as desired and confirm.

4.20 Image source selection

About image source selection

By default, your Nio Fusion 12MP automatically detects the connected image sources and attaches them to
the correct side of the screen. You can also manually select the image sources and control on which side of
the screen they must be displayed.

The

Image source selection

menu is not available when the

KVM switch

is enabled (to disable, see

“KVM switch”, page 37

).

To automatically select image sources

1.

Bring up the OSD main menu.

2.

Navigate to the

Configuration > Image Sources

menu.

3.

Enter the

Image Source Selection

submenu.

4.

Select the

Automatic

image source selection mode and confirm.

To manually select image sources

1.

Bring up the OSD main menu.

2.

Navigate to the

Configuration > Image Sources

menu.

3.

Enter the

Image Source Selection

submenu.

4.

Select one of the following image source selection modes and confirm:

One Image Source:

Allows to select one image source to display.

Two Image Sources:

Displays DisplayPort IN 1 on the left side and DisplayPort IN 2 on the right side

of the screen.

Expert mode:

Allows to select for the left and right side of the screen which image source to display:

DisplayPort IN 1, DisplayPort IN 2 or None.

4.21 KVM switch

About the KVM switch

The Nio Fusion 12MP can be used in dual workstation setup with KVM switch. This allows the display to be
connected to two different workstations and control both with a single keyboard and mouse connected to the
display. Switching the video and controls signals between the two workstations is done with the KVM switch,
which must be enabled as explained below.

Your display system must be specifically set up and configured for KVM input switching. For
instructions, see

“Dual workstation setup with KVM switch”, page 13

To enable/disable the KVM switch

1.

Bring up the OSD main menu.

2.

Navigate to the

Configuration > Image Sources

menu.

Advanced operation

Содержание Nio Fusion 12MP

Страница 1: ...ENABLING BRIGHT OUTCOMES MDNC 12130 User guide Nio Fusion 12MP...

Страница 2: ...Registered office Barco NV President Kennedypark 35 8500 Kortrijk Belgium www barco com en support www barco com Barco NV Beneluxpark 21 8500 Kortrijk Belgium www barco com en support www barco com...

Страница 3: ...operation 22 3 2 Shortcut bar 22 3 3 Standby switching 23 3 4 SoftGlow light mode switching 23 3 5 KVM input switching 24 3 6 OSD menu access and use 24 4 Advanced operation 25 4 1 OSD menu language 2...

Страница 4: ...4 22 Grayscale conversion modes 38 4 23 DisplayPort standard version 38 4 24 EDID format 39 4 25 EDID timings 39 4 26 Display info 40 4 27 Display status 40 4 28 SteadyColor 40 4 29 Maximum luminance...

Страница 5: ...5 R5913350 05 Nio Fusion 12MP 1 1 What s in the box 6 1 2 At a glance 6 1 3 Compatible Barco system components 9 Welcome 1...

Страница 6: ...em sheet Cables for Mains Video and USB If you ordered a Barco MXRT display controller it is also in the box together with its accessories A dedicated user guide is available on the documentation disc...

Страница 7: ...l Press Activate the shortcut bar go into sub menus confirm adjustments and selections Turn left right Scroll through sub menus change values make selections Press and hold for approximately 3 seconds...

Страница 8: ...tch enabled 14 DisplayPort IN 1 also Main for KVM switch 15 DisplayPort IN 2 also Secondary for KVM switch 16 Kensington security slot 17 100 240 VAC mains power input and power switch Display positio...

Страница 9: ...erview Following Barco components are compatible with your Nio Fusion 12MP display QAWeb Enterprise version 2 5 or later or MediCal QAWeb version 1 13 20 or later MXRT x700 range or later display cont...

Страница 10: ...R5913350 05 Nio Fusion 12MP 10 Welcome...

Страница 11: ...Fusion 12MP 2 1 Single workstation setup 12 2 2 Dual workstation setup with KVM switch 13 2 3 QAWeb registration 16 2 4 VESA mount installation 16 2 5 VESA mount installation in portrait mode 18 Insta...

Страница 12: ...r guide on the included documentation CD or at www barco com support Connect the cables 1 Remove the connector cover 2 Connect the DisplayPort inputs DP IN 1 and DP IN 2 with the DisplayPort outputs o...

Страница 13: ...mxrt 2 Install the MXRT driver and Intuitive Workflow Tools as described in the Barco Display Controller and Intuitive Workflow Tools user guide on the included documentation CD or at www barco com su...

Страница 14: ...host on workstation 1 4 Connect DisplayPort input 2 DP IN 2 with the DisplayPort output on workstation 2 5 Connect the secondary USB upstream connector USB UP 2 with a USB host on workstation 2 Note T...

Страница 15: ...one DisplayPort video cable 1 Bring up the OSD main menu 2 Navigate to Configuration Image Sources Input Settings EDID format 3 Select DisplayID 1 3 and confirm 4 Navigate to Configuration Image Sourc...

Страница 16: ...n still using the older MediCal QAWeb visit www barco com support medical qaweb to obtain the installation package and user documentation 2 4 VESA mount installation CAUTION Use suitable mounting appa...

Страница 17: ...een 3 Push the VESA mount cover to the right and lift the cover on the right with your finger or a flathead screwdriver Image 2 6 4 Push the VESA mount cover to the left and release the cover on the l...

Страница 18: ...5 mm Lmax T W 10 mm T W Lmin Lmax Image 2 10 2 5 VESA mount installation in portrait mode Introduction By default your Nio Fusion 12MP display is mounted in landscape position but it can also be used...

Страница 19: ...able can be connected This will give you the full resolution at reduced refresh rate but with just one Windows taskbar at the bottom of the screen 3 For your convenience align the display s OSD menu o...

Страница 20: ...R5913350 05 Nio Fusion 12MP 20 Installation and setup...

Страница 21: ...io Fusion 12MP 3 1 Recommendations for daily operation 22 3 2 Shortcut bar 22 3 3 Standby switching 23 3 4 SoftGlow light mode switching 23 3 5 KVM input switching 24 3 6 OSD menu access and use 24 Da...

Страница 22: ...regularly to avoid image retention of the static elements Understand pixel technology LCD displays use technology based on pixels As a normal tolerance in the manufacturing of the LCD a limited number...

Страница 23: ...ow light mode switching About the SoftGlow lights The SoftGlow Task light sheds a light on the desktop while the SoftGlow Wall light provides ambient lighting for the reading room to reduce eye fatigu...

Страница 24: ...is selected by default when the KVM switch is enabled When switching KVM inputs the video is switched instantly while switching the USB signal might take a little longer up to 2 seconds 3 6 OSD menu a...

Страница 25: ...4 4 18 Embedded QA 34 4 19 Image scaling 36 4 20 Image source selection 37 4 21 KVM switch 37 4 22 Grayscale conversion modes 38 4 23 DisplayPort standard version 38 4 24 EDID format 39 4 25 EDID timi...

Страница 26: ...R5913350 05 Nio Fusion 12MP 26 Advanced operation...

Страница 27: ...menu 4 Select Landscape Portrait or Inverse Portrait and confirm 4 3 OSD menu automatic close function About the OSD menu automatic close function By default the OSD menu will disappear automatically...

Страница 28: ...support To adjust the SoftGlow Task light brightness 1 Bring up the OSD main menu 2 Navigate to the Configuration Lights menu 3 Enter the SoftGlow Task Light brightness submenu 4 Set a SoftGlow Task L...

Страница 29: ...y but it also needs to be activated on your workstation To do this go to the Power options of your workstation Barco recommends setting DPMS activation after 20 minutes of non usage When DPMS mode is...

Страница 30: ...e target 1 Bring up the OSD main menu 2 Navigate to the Configuration Calibration menu 3 Enter the Luminance Target submenu 4 Set a luminance target value as desired and confirm The default factory ca...

Страница 31: ...vigate to the Configuration Calibration Color Settings Color Presets menu 3 Select User and confirm 4 Enter the Color Definition submenu select Color Coordinates and confirm 5 Enter the x and y submen...

Страница 32: ...splay are Native The native display panel behavior will not be corrected Dynamic Gamma 1 8 or 2 2 These are gamma functions that are shifted to take into account the non zero luminance of an LCD panel...

Страница 33: ...m Therefore it is also important to select a realistic reading room when enabling ALC This can be done by following the instructions in Reading rooms page 33 To enable disable ALC 1 Bring up the OSD m...

Страница 34: ...the integrated ambient light sensor into account Continuous ALC can only be enabled on your display when the display function is set to DICOM Please refer to Display functions page 32 To select contin...

Страница 35: ...ance Black luminance Shows the measured black luminance DICOM Calibration Status No calibration executed yet No other information is visible Calibration executed When the calibration is executed the f...

Страница 36: ...enu 3 Enter the DICOM Preferences submenu 4 Select Reset DICOM Calibration to restore the original not corrected DICOM curve 4 18 6 DICOM error threshold About DICOM error threshold The threshold to d...

Страница 37: ...guration Image Sources menu 3 Enter the Image Source Selection submenu 4 Select one of the following image source selection modes and confirm One Image Source Allows to select one image source to disp...

Страница 38: ...ode is possible but then your display s image source selection mode should be set to allow this Refer to Image source selection page 37 to do this About grayscale conversion modes Grayscale conversion...

Страница 39: ...h KVM switch page 13 To set the EDID format 1 Bring up the OSD main menu 2 Navigate to the Configuration Image Sources Input Settings DisplayPort 1 2 menu 3 Enter the EDID format submenu 4 Select one...

Страница 40: ...earity in the display of color images With multiple internal sensors the display parameters are constantly monitored and adjusted to ensure that the perceptual difference between colors is proportiona...

Страница 41: ...Service menu 5 Navigate to the Maximum Luminance demo submenu and confirm The maximum luminance demo is activated 6 Press the jog dial to manually cancel the maximum luminance demo or wait for 60 seco...

Страница 42: ...R5913350 05 Nio Fusion 12MP 42 Advanced operation...

Страница 43: ...43 R5913350 05 Nio Fusion 12MP Cleaning the display 5...

Страница 44: ...ut a certain cleaning product use plain water Do not use any of the following products Alcohol in concentrations 70 Strong alkalis lye strong solvents Acetone Toluene Acids Detergents containing fluor...

Страница 45: ...45 R5913350 05 Nio Fusion 12MP Repackaging instructions 6...

Страница 46: ...ut the display in the lowest position and lock the stand locking mechanism Image 6 1 2 Place the display in the bottom buffer on a stable surface Caution It takes 2 persons to safely execute this acti...

Страница 47: ...47 R5913350 05 Nio Fusion 12MP Image 6 3 4 Position the front buffer Image 6 4 5 Slide the accessory box es and all cables in the dedicated cavities Repackaging instructions...

Страница 48: ...R5913350 05 Nio Fusion 12MP 48 Image 6 5 6 Slide the display and the accessories in the box Image 6 6 7 Close and seal the box Repackaging instructions...

Страница 49: ...49 R5913350 05 Nio Fusion 12MP Image 6 7 Repackaging instructions...

Страница 50: ...R5913350 05 Nio Fusion 12MP 50 Repackaging instructions...

Страница 51: ...2 Cybersecurity 53 7 3 Environmental information 54 7 4 Regulatory compliance information 55 7 5 EMC notice 57 7 6 Explanation of symbols 59 7 7 Legal disclaimer 63 7 8 Open source license information...

Страница 52: ...part The equipment shall not be used with life support equipment The user should not touch the equipment nor its signal input ports SIP signal output ports SOP and the patient at the same time Power...

Страница 53: ...f the power cord or plug is damaged or frayed If liquid has been spilled into the equipment If the equipment has been exposed to rain or water If the equipment does not operate normally when the opera...

Страница 54: ...details please visit the Barco website at http www barco com AboutBarco weee Turkey RoHS compliance T rkiye Cumhuriyeti AEEE Y netmeli ine Uygundur Republic of Turkey In conformity with the WEEE Regu...

Страница 55: ...and marked with the Environmental Friendly Use Period EFUP logo The number inside the EFUP logo that Barco uses please refer to the photo is based on the General guidelines of environment friendly use...

Страница 56: ...e received including interference that may cause undesired operation This device has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules T...

Страница 57: ...s CISPR 11 Group 1 The Nio Fusion 12MP uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...

Страница 58: ...ruptions 0 residual voltage for 250 periods at 0 0 residual voltage for 0 5 period at 0 45 90 135 180 225 270 and 315 0 residual voltage for 1 period at 0 70 residual voltage for 25 periods at 0 Volta...

Страница 59: ...2 0 3 28 5240 5100 5800 W LAN 802 11 a n Pulse modulation 217 Hz 0 2 0 3 9 5500 5785 7 6 Explanation of symbols Symbols on the device On the device or power supply you may find the following symbols n...

Страница 60: ...SE regulations Indicates the device is approved according to the RCM regulations Indicates the device is approved according to the EAC regulations Caution Federal law United Stated of America restrict...

Страница 61: ...ndicates the Authorised Representative for the European Union REP CH Indicates the Authorised Representative for Switzerland Warning dangerous voltage Caution Consult the Instructions For Use eIFU ind...

Страница 62: ...m number of identical boxes which may be stacked on each other where n is the limiting number xx yy Kg or xx yy xx yy Indicates the weight of the box and that it should be carried with two persons Ind...

Страница 63: ...se process will be triggered immediately To address specific security concerns or to report security issues with Barco products please inform us via contact details mentioned on https www barco com ps...

Страница 64: ...arco will charge you for the actual costs of performing such distribution such as the cost of media shipping and handling You may exercise this option upon written request to Barco N V attn legal depa...

Страница 65: ...25 Swivel 30 to 30 Pivot N A Height adjustment range 100 mm Mounting standard VESA 100 mm Screen protection N A Recommended modalities All digital images including digital mammography and breast tomos...

Страница 66: ...deo cables USB cables Mains cables Optional accessories Display controller QA software QAWeb Warranty 5 years including 40000 hrs backlight warranty Operating temperature 0 C to 35 C 20 C to 30 C with...

Страница 67: ...67 R5913350 05 Nio Fusion 12MP Relation of storage and operating humidity and temperature Image 7 1 Important information...

Страница 68: ...R5913350 05 Nio Fusion 12MP 68 Important information...

Страница 69: ......

Страница 70: ...www barco com R5913350 05 2022 12 19 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...

Отзывы: