User manual MFGD 1318
29
Low Pass Filter*
* Not in DVI mode
The display electronics provide 4 different low-pass filters to filter
possible high-frequent distortion from an
analog
video signal.
Select a different filter in case the image contains high-frequent noise
or distortion. Select the filter that gives the best result.
Readjustment of sample phase is necessary after selection of a
different filter.
4.3.
Advanced functions in the Settings menu
The Settings menu contains the following advanced functions:
Display function
The display contains a number of factory-defined and user-defined
Display functions (LUTs).
Turn the control wheel to select a display function.
- Select a
DICOM
display function for most medical viewing
applications. The DICOM function results in more visible grayscales
in the images.
- Select a
CRT
(gamma) display function in case the display is used
to replace a CRT display.
- LUT 3 and LUT 4 are user-programmable display functions
(dedicated software needed).
- If you select
None
, no display function will be selected.
- The selection
Test LUT
is meant for internal test purposes. It should
not be selected for normal operation.
Input Mode
To change the Input Mode setting, click the control wheel. Turn the
wheel to select Standard or Extended input mode. Then click the wheel
again to enter the selected submenu.
The following input mode settings can be selected:
Standard input mode
RGB->Y ................... The RGB video signals from the imaging board
are calculated and transformed into a single
luminance value according to the formula
0.3R + 0.59G + 0.11B. This single luminance
signal drives the RGB sub-pixels of the panel
equally.