7. Important information

Immunity test

IEC 60601

Test levels

Compliance level

Electromagnetic
environment –
guidance

separation distance in
meters (m).

Field strengths from

fi

xed RF transmitters,

as determined by an
electromagnetic site
survey,

3

should be less

than the compliance
level in each frequency
range.

4

Interference may occur in
the vicinity of equipment
marked with symbol:

At 80 MHz and 800 MHz, the higher frequency range applies.

These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and re

fl

ection from structures, objects and people.

Recommended separation distance

The Eonis is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the Eonis can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Eonis as recommended below, according to the maximum output power of the communications
equipment.

Separation distance according to frequency of transmitter

Rated maximum output
power of transmitter

5

W

150kHz to 80MHz

d=1.2

P

80MHz to 800MHz

d=1.2

P

800MHz to 2.5GHz

d=2.3

P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the

location in which the Eonis is used exceeds the applicable RF compliance level above, the Eonis should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Eonis.

4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

5. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be

estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the

transmitter in watts (W) according to the transmitter manufacturer.

30

K5903081 EONIS 19/09/2014

Содержание MDRC-2224 WP

Страница 1: ...Eonis User Guide 24 inch clinical display MDRC 2224 WP MDRC 2224 BL K5903081 03 19 09 2014...

Страница 2: ...Barco nv President Kennedypark 35 8500 Kortrijk Belgium Phone 32 56 23 32 11 Fax 32 56 26 22 62 Support www barco com esupport Visit us at the web www barco com Printed in Belgium...

Страница 3: ...hrough the OSD menus 14 5 Advanced operation 15 5 1 Video input source selection 15 5 2 Luminance adjustment 15 5 3 Gamma selection 15 5 4 Ambient light reading room selection 16 5 5 White point selec...

Страница 4: ...Table of contents 2 K5903081 EONIS 19 09 2014...

Страница 5: ...required in clinical environments It features an inte grated cable management system making it safe to use in hospitals and private practices The multi ple mounting options and connections further add...

Страница 6: ...ning MediCal QAWeb Agent a set of AC power cord s depending on the region of operation a VGA cable a HDMI cable a DP cable a USB cable an external power supply an accessory bag cleaning cloth velcro c...

Страница 7: ...rs 2 PARTS CONTROLS AND CONNECTORS 2 1 Front view Overview Image 2 1 1 Front sensor 2 USB downstream connector 3 Left Decrease key 4 Right Increase key 5 Menu Enter key 6 Standby key 7 Power status LE...

Страница 8: ...2 1 Connector compartment cover 2 Kensington security slots 3 Opening for cable routing strap 4 12 VDC power input 5 USB downstream connector 6 USB upstream connector 7 VGA input 8 HDMI input 9 Displa...

Страница 9: ...orientation of the video input via the screen settings of the computer WARNING Put the display in its highest position before pivoting the display The display will be delivered in portrait mode Pivot...

Страница 10: ...th the display s USB upstream connector by means of the supplied USB cable 3 Connect a speaker amplifier with the display s audio line out if you want to use the internal sound option of your display...

Страница 11: ...ugh the opening in the back of the stand 2 Bundle all cables together so that they will fit in the strap 3 Wrap and fix the cable routing strap around all cables Image 3 5 3 6 Kensington security slot...

Страница 12: ...m Please make sure that the length of the screws is 10 mm VESA plate thickness tolerance of 0 5 mm Image 3 7 CAUTION Use an arm that can support a weight of at least 12 kg 26 50 lbs Failure to do so c...

Страница 13: ...3 Display installation CAUTION Use suitable mounting apparatus to avoid risk of injury K5903081 EONIS 19 09 2014 11...

Страница 14: ...3 Display installation 12 K5903081 EONIS 19 09 2014...

Страница 15: ...sage A good screen saver displays moving content In case you are working with the same image or an application with static image elements for several hours continuously so that the screen saver is not...

Страница 16: ...settings through the OSD menu To bring up the OSD menu 1 While the display is switched on press the Menu Enter key As a result the OSD main menu comes up in the middle of the screen If no further act...

Страница 17: ...nu 2 Navigate to the Input Selection menu 3 Select one of the available video input sources 5 2 Luminance adjustment About luminance adjustment The luminance of your Eonis display is adjustable over a...

Страница 18: ...light reading room selection About ambient light reading rooms The available ambient light reading rooms for your display are Dark Room Corresponds to light conditions in dark diagnostic reading room...

Страница 19: ...of the analog video clock frequency clock phase horizontal position vertical position Level Allows to manually adjust the contrast and brightness levels of the analog video To adjust the analog video...

Страница 20: ...Navigate to the Adjustments Settings menu 3 Enter the Power Status LED submenu 4 Change the behavior of the power status LED as desired and confirm The orange standby power saving state of the LED is...

Страница 21: ...selected when the self calibration frequency is defined by MediCal QAWeb To adjust the self calibration frequency 1 Bring up the OSD main menu 2 Navigate to the Adjustments Settings menu 3 Enter the S...

Страница 22: ...5 Advanced operation 20 K5903081 EONIS 19 09 2014...

Страница 23: ...the cleaning product In case of doubt about a certain cleaning product use plain water Do not use following products For black model alcohol solvents at higher concentration 5 for white model alcohol...

Страница 24: ...6 Maintenance 22 K5903081 EONIS 19 09 2014...

Страница 25: ...oxide Non patient care equipment Equipment primarily for use in a health care facility that is intended for use where contact with a patient is unlikely no applied part The equipment may not be used...

Страница 26: ...nment as defined in the standard as at least 1 5 meters from the patient or the patient support Any person who connects external equipment to signal input signal output or other connectors has formed...

Страница 27: ...this product please contact your local city office or your municipal waste disposal service For details please visit the Barco website at http www barco com en AboutBarco weee Turkey RoHS compliance T...

Страница 28: ...the homogeneous materials used for this part is above the limit requirement in SJ T11363 2006 EIP EFUP Barco EFUP All Electronic Information Products EIP that are sold within Chinese Mainland must com...

Страница 29: ...device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This device generates uses and can rad...

Страница 30: ...te medical EMC standards on emissions to and interference from sur rounding equipment Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this d...

Страница 31: ...ital environment If the user of the Eonis requires continued operation during power mains interruptions it is recommended that the Eonis be powered from an uninterruptible power supply or a battery Po...

Страница 32: ...Rated maximum output power of transmitter 5 W 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23...

Страница 33: ...iance with the Directive 93 42 EEC as Class II device Indicates compliance with Part 15 of the FCC rules Class A or Class B Indicates the device is approved according to the UL regulations Indicates t...

Страница 34: ...nic Equipment directive Indicates Direct Current DC Indicates Alternating Current AC Stand by Equipotentiality Symbols on the box On the box of the device you may find the following symbols nonrestric...

Страница 35: ...ou with the most accurate and usable documentation possible if you discover errors please let us know Barco software products are the property of Barco They are distributed under copyright by Barco N...

Страница 36: ...l Housing color Black Silver Video input signals VGA DisplayPort HDMI USB ports 1 upstream 2 downstream USB standard 2 0 Power requirements nominal 100 240 V Power consumption nominal 40 W 1 0 W in st...

Страница 37: ...anty 3 years Operating temperature 10 C 35 C Storage transport temperature 20 C 60 C Operating humidity 10 90 non condensing Storage transport humidity 5 90 non condensing at 39 deg ambient Operation...

Страница 38: ...lass I product CE 2004 108 EC 2014 30 EU Directive 2004 108 EC is repealed with effect from 20 April 2016 IEC 60601 1 ED 3 ANSI AAMI ES60601 1 2005 C1 09 A2 10 CAN CSA C22 2 No 60601 1 08 DEMKO EN 606...

Страница 39: ...7 Important information Operating pressure 70 kPa 106 kPa Storage transport pressure 50 kPa 106 kPa K5903081 EONIS 19 09 2014 37...

Страница 40: ...7 Important information 38 K5903081 EONIS 19 09 2014...

Страница 41: ...adjustment Check and re configure the display mode of the vertical refresh rate of your graphics card to make it compatible with the display Problem Image is unstable and flickering Use Phase to make...

Страница 42: ...ans that the signal of the computer graphics card is not compatible with the display When the signal is not included in the compatibility mode we have listed in the appendices of this manual the displ...

Отзывы: