5. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the Nio Color 2MP
23" requires continued
operation during power
mains interruptions, it
is recommended that
the Nio Color 2MP 23"
be powered from an
uninterruptible power
supply or a battery.
Power frequency (47-63
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
2
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Nio Color 2MP
23", including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800
MHz
d = 2.3
√
P
800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
fi
xed RF transmitters,
as determined by an
electromagnetic site
survey,
3
should be less
1. is the a.c. mains voltage prior to application of the test level.
2. Nio Color 2MP 23" doesn’t contain susceptible components to magnetic fields
K5902134 NIO COLOR 2MP 23" 05/07/2017
27
Содержание MDNC-2123
Страница 1: ...Nio Color 2MP 23 User Guide MDNC 2123 option DE K5902134 01 05 07 2017...
Страница 4: ...Table of contents 2 K5902134 NIO COLOR 2MP 23 05 07 2017...
Страница 14: ...2 Installation 12 K5902134 NIO COLOR 2MP 23 05 07 2017...
Страница 20: ...3 Operation 18 K5902134 NIO COLOR 2MP 23 05 07 2017...