29
R5911227 /05
Coronis Fusion 4MP / 6MP
To select a reading room:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Reading Room
submenu.
4.
Select one of the available reading rooms and confirm.
4.17 Continuous Ambient Light Compensation
(ALC)
About Continuous ALC
Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged
ambient light, as measured by the integrated ambient light sensor, into account.
Continuous ALC can only be enabled on your display when the display function is set to DICOM.
Please refer to
.
To select continuous ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Continuous ALC
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.18 Embedded QA
4.18.1 About Embedded QA
About
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display’s OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or QAWeb?
Embedded QA is not a replacement for the Barco QAWeb solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends QAWeb as the solution of choice for calibration and QA. QAWeb brings many benefits
such as centralized asset management, the ability to schedule tasks, remote management, automated
reporting, alerting and specific support of regional QA standards such as DIN 6868-57, JESRA and AAPM
TG18. That's why QAWeb Agent acts as the master for all supported displays from the moment it is installed
and running. QAWeb Agent will take over from Embedded QA and overwrite any settings which were applied
by Embedded QA.
4.18.2 DICOM status report
About DICOM status report
Following information is available:
DICOM Compliance Status
(status since last compliance check)
•
Compliance status:
Shows if the current DICOM curve is compliant or not.
Содержание MDCC-4430
Страница 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 NC MDCC 6530 NC User guide Coronis Fusion 4MP 6MP...
Страница 5: ...5 R5911227 05 Coronis Fusion 4MP 6MP Welcome 1...
Страница 8: ...R5911227 05 Coronis Fusion 4MP 6MP 8 Welcome...
Страница 9: ...9 R5911227 05 Coronis Fusion 4MP 6MP Installation 2...
Страница 16: ...R5911227 05 Coronis Fusion 4MP 6MP 16 Installation...
Страница 17: ...17 R5911227 05 Coronis Fusion 4MP 6MP Daily operation 3...
Страница 35: ...35 R5911227 05 Coronis Fusion 4MP 6MP Cleaning your display 5...
Страница 37: ...37 R5911227 05 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Страница 41: ...41 R5911227 05 Coronis Fusion 4MP 6MP Image 6 7 Repackaging instructions...
Страница 42: ...R5911227 05 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Страница 43: ...43 R5911227 05 Coronis Fusion 4MP 6MP Important information 7...
Страница 62: ...R5911227 05 Coronis Fusion 4MP 6MP 62 Important information...
Страница 63: ......
Страница 64: ...www barco com R5911227 05 2022 01 12 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...