3. Display installation
Protective earth
Earth the MDSC-2326 by connecting the protective earth pin to a grounded outlet by means of a yel-
low/green AWG18 wire (maximum admitted cable length according to national regulation requirements).
1
Image 3-8
C
AUTION
:
The display must be earthed.
Potential equalization
When potential equalization between the display and other devices is required then connect the potential
equalization pin (POAG: DIN42801) to the potential equalization terminal of the equipment.
3.7
Cable routing
To route the cables
1. For displays mounted to a VESA arm with internal cable routing provisions, route all cables through the
cable routing channel, then reinstall the connector compartment cover.
8a
Image 3-9
20
K5902135 (451920612562) MDSC-2326 01/12/2017