4.13 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been defined. These functions emphasize on these parts containing crucial diagnostic
information by correcting the native panel behavior.
The available display functions for your display are:
•
Native:
The native panel behavior will not be corrected.
•
Dynamic Gamma 1.8
or
2.2
: These are gamma functions that are shifted to take into account the non-zero
luminance of an LCD panel when driven with a
“
black
”
signal. They are especially useful in CT applications
to improve the perception of low Hounsfield values.
•
DICOM:
DICOM (Digital Imaging and Communications in Medicine) is an international standard that was
developed to improve the quality and communication of digital images in radiology. In short, the DICOM
display function results in more visible grayscales in the images. Barco recommends selecting the DICOM
display function for most medical viewing applications.
•
sRGB:
The sRGB color space combines a display function, color space and white point selection, and is
designed to match typical home and office viewing conditions. It is widely used in most computer
applications.
Note:
This setting does not fully set your display to the sRGB color space. Refer to
”
,
page 27 to do this.
•
User:
This display function will be automatically selected when display functions are defined by MediCal
QAWeb.
•
Gamma 1.8
or
2.2
: Select one of these display functions in case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
•
Test:
For Barco service purposes only.
The settings of the display must be adapted to suit the requirements of the visualization software. In
case of doubt, please contact the vendor of the visualization software.
To select a display function:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Enter the
Display Function
submenu.
4.
Select one of the available display functions and confirm.
4.14 sRGB
About sRGB
When sRGB is enabled, the display is set to an sRGB color space with the following settings:
•
Color preset:
6500K. See
, page 25 for more info.
•
Color space:
sRGB. See
, page 26 for more info.
•
Display function:
sRGB. See
, page 27 for more info.
To enable sRGB:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Select
sRGB
and confirm.
Содержание Coronis Fusion 4MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 MDCC 6530 User guide Coronis Fusion 4MP 6MP...
Страница 5: ...5 R5911227 02 Coronis Fusion 4MP 6MP Welcome 1...
Страница 8: ...R5911227 02 Coronis Fusion 4MP 6MP 8 Welcome...
Страница 16: ...R5911227 02 Coronis Fusion 4MP 6MP 16 Installation...
Страница 17: ...17 R5911227 02 Coronis Fusion 4MP 6MP Daily operation 3...
Страница 21: ...21 R5911227 02 Coronis Fusion 4MP 6MP Advanced operation 4...
Страница 34: ...R5911227 02 Coronis Fusion 4MP 6MP 34 Advanced operation...
Страница 35: ...35 R5911227 02 Coronis Fusion 4MP 6MP Cleaning your display 5...
Страница 37: ...37 R5911227 02 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Страница 41: ...R5911227 02 Coronis Fusion 4MP 6MP 41 Image 6 7 Repackaging instructions...
Страница 42: ...R5911227 02 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Страница 43: ...43 R5911227 02 Coronis Fusion 4MP 6MP Important information 7...
Страница 59: ...R5911227 02 Coronis Fusion 4MP 6MP 59 Important information...