7.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Type of protection (electrical)
Monitor with internal power supply: Class I equipment.
Degree of safety (flammable anesthetic mixture)
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Non-patient care equipment
•
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
•
The equipment shall not be used with life support equipment.
•
The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the
patient at the same time.
Power connection
–
Equipment with internal power supply
•
This equipment must be earthed.
•
Power requirements: The equipment must be powered by the AC mains voltage.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-
voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
High magnetic environment
•
The device shall not be used in the high magnetic environment of an MRI scanner.
•
The installer shall assess the magnetic environment before installation or use of the device.
Power cords:
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to be
walked upon or pinched by items placed upon or against them, paying particular attention to cords at plugs
and receptacles.
Содержание Coronis Fusion 4MP
Страница 1: ...ENABLING BRIGHT OUTCOMES MDCC 4430 MDCC 6530 User guide Coronis Fusion 4MP 6MP...
Страница 5: ...5 R5911227 02 Coronis Fusion 4MP 6MP Welcome 1...
Страница 8: ...R5911227 02 Coronis Fusion 4MP 6MP 8 Welcome...
Страница 16: ...R5911227 02 Coronis Fusion 4MP 6MP 16 Installation...
Страница 17: ...17 R5911227 02 Coronis Fusion 4MP 6MP Daily operation 3...
Страница 21: ...21 R5911227 02 Coronis Fusion 4MP 6MP Advanced operation 4...
Страница 34: ...R5911227 02 Coronis Fusion 4MP 6MP 34 Advanced operation...
Страница 35: ...35 R5911227 02 Coronis Fusion 4MP 6MP Cleaning your display 5...
Страница 37: ...37 R5911227 02 Coronis Fusion 4MP 6MP Repackaging instructions 6...
Страница 41: ...R5911227 02 Coronis Fusion 4MP 6MP 41 Image 6 7 Repackaging instructions...
Страница 42: ...R5911227 02 Coronis Fusion 4MP 6MP 42 Repackaging instructions...
Страница 43: ...43 R5911227 02 Coronis Fusion 4MP 6MP Important information 7...
Страница 59: ...R5911227 02 Coronis Fusion 4MP 6MP 59 Important information...