28
User’s Guide
Electromagnetic Compatibility
Like other electrical medical equipment, the
AMM240ED
requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the AMM240ED must be installed and operated
according to the EMC information provided in this manual.
Note
The AMM240ED has been designed and tested to comply with IEC 60601-1-
2:2014 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the AMM240ED.
Warning
Do not use cables or accessories other than those provided
with the AMM240ED, as this may result in increased electromagnetic emis-
sions or decreased immunity to such emissions.
Warning If the AMM240ED is used adjacent to or stacked with other
equipment, observe and verify normal operation of the AMM240ED in the
a surgical procedure.
Consult the tables below for guidance in placing the AMM240ED.
The customer or the user of AMM240ED should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Model AMM240ED uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
RF emissions CISPR 11
Class B
AMM240ED is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations
IEC61000-3-3
Complies
Manufacturer’s declaration - electromagnetic emission
The customer or the user of AMM240ED should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Floors should be wood,
concrete or ceramic tile.
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient
/burst IEC61000-4-4
± 2 kV
100 kHz repetition frequency
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
± 0,5 kV, ± 1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment.
Manufacturer’s Declaration - Electromagnetic Immunity