Specifications

Instrument
Measuring apertures:  

reflection 2 and 8mm (adjustable)
transmission 2 and 8mm (adjustable)
(Model RT additional 6mm)

Geometry: 

reflection 45°:0° circumferential
transmission d:0

Calibration:

automatic with internal white reference

Physical illumination:

type A

Measurement time:

< 0.3 seconds

Measuring speed:

approx. 10 min for 1248 fields

Repeatability:

< 0.2 

∆Eab 94 on white or

±

0.005 D (up to 1.000 D)

Inter-instrument  agree-
ment:

typ 1.0 

∆Eab 94

Measuring sensor: 

diffraction grating with diode array

Spectral resolution: 

3.5 nm

Spectral range: 

380 ... 780 nm

Density measuring range: 0 ... 2.3 D 

Target thickness 

max 20mm 

Maximum measuring area:  290 x 200mm (LxH)
Smallest measuring step: 0.2mm
Interface:

USB, serial 57600 Baud for PC and MAC

Dimensions:

571 x 433 x 160mm (LxWxH)

Weight:

11 kg net

Specifications are subject to change without notice

Содержание Spectro LFP S3

Страница 1: ...Operating Manual Spectro LFP S3 RT and Tex automatic reflection and transmission spectrophotometer Edition E5...

Страница 2: ...2 Spectro LFP RT Operating Manual...

Страница 3: ...e samples 15 3 3 Measuring special materials 16 3 3 1 Measuring heavy reflective materials 16 3 3 2 Measuring thick transparent materials 17 3 3 3 Measuring special transparent materials 17 Chapter 4...

Страница 4: ...r use in color management for the digital digital imaging market Maywesuggestyoureadtheoperatingmanualfirstforfullyenjoyingthe capabilities of this instrument thus you will be able to get familiar wit...

Страница 5: ...be measured automatically papers vinyl textiles cardboards plastic plates gypsum plates wood glass stone and ceramics plates For LFP market Individual ICC profiles for large format output equipment p...

Страница 6: ...software for Mac OSX and Windows power supply adapter this operating manual package keep it for possible transports 1 3 Components of the Spectro LFP Tex instrument The Spectro LFP Tex comes with the...

Страница 7: ...sample holders for special materials code C200A50 S1 consisting of 1 Reflectionsampleholderforextraheavy reflective materials code C200A50 2 Transmission sample holder for thick transparent materials...

Страница 8: ...1 Installing on a Mac computer under OS X Switch off the Spectro LFP Insert the supplied CD Rom into your computer run the Barbieri Gateway software installer or the USB Driver Installer program on t...

Страница 9: ...driver Switch off the Spectro LFP Insert the supplied CD Rom into your computer run the Barbieri Gateway software installer or the USB Driver Setup program on the CD and follow the instructions The in...

Страница 10: ...hotometer is connected to the rear side of the instrument When connecting the instrument to an Apple computer use the USB cable connection WhenconnectingtoaWindowsPC youhavethechoicebetweenserial or U...

Страница 11: ...ment with Barbieri electronic in order to get access to privileged information like access to download area product information firmware updates etc Registration can be done through the Barbieri Gatew...

Страница 12: ...pectroLFPneartheswitchwilllight The instrument performs the following movements the measuring head goes in its up position the xy table moves to the right side near the measuring head and moves then t...

Страница 13: ...of approx 97 C value 2 Measurements made with this sample holder are also named white backing measurements as recommended by the ICC International Color Consortium for the creation of ICC profiles If...

Страница 14: ...The sample to be measured must be fixed inside the indicated corner marks using a tape When fixing the sample make sure the target is printed in a straight position This may be important particularly...

Страница 15: ...ing transmissive samples Transmissive materials must be measured using the supplied trans mission sample holder model 3 For measuring insert the sample between the two transparent sheets of the sample...

Страница 16: ...ctivematerials theoptionalsampleholder for heavy materials must be used It consists of a metallic holder with 2 magnets which are used to fix the sample The anti slip mat helps keep target from moving...

Страница 17: ...e larger must the patches be For material of 5mm thickness the recommended patch size is 10mm The sample is fixed to the sample holder using a tape 3 3 3 Measuring special transparent materials When y...

Страница 18: ...ard placed under the measuring head Note The validity of this reference white is 24 months from the pro duction date See the serial no on the reference white The first four charactersindicatetheyearan...

Страница 19: ...hite calibration Nearly all linearization and profiling targets require to calibrate against the white of the material to be measured assuming this as the refer ence white with a value of L 100 a 0 b...

Страница 20: ...of the target including its white point As the L value can be as low as 50 or lower for diffuse materials this method is not recommended for linearization and color management applications as it will...

Страница 21: ...fore the auto position line is visible and not covered by the sample holder or tape 4 3 Measuring modes The Spectro LFP RT supports three measuring modes 1 Up Down mode In up down mode the measuring h...

Страница 22: ...ture the wide aperture to 8mm Pressing the button will select the following apertures LED off the 2mm aperture is selected LED on intermittent the 6mm aperture is selected LED on always 8mm aperture i...

Страница 23: ...to be measured is determined in only 1 position it is important that the materialhasthesamethicknessonthewholemeasur ing area Tolerances of up to 1 mm are acceptable max 2mm The instrument supports me...

Страница 24: ...ilter cover The cover is holding by a magnet The UV cut filter is used to avoid wrong measurements when measur ing samples with optical brighteners Measurements correspond to measurement condition M2...

Страница 25: ...in text or XML format The software runs under either Windows or Mac OSX and delivers measurement data either spectral CIELab or density The software is available for download at the following web site...

Страница 26: ...n standard Replace the reference white of the Spectro LFP RT if it is no more valid or it is somehow damaged Any scratches dust or dirt on the surface of the reference white may affect the accuracy of...

Страница 27: ...e a measuring lamp of Spectro LFP RT if it is obvious that a lamp is broken Reflection When the instrument is switched on the reflection lamps are blinking Check if you see the light of all the three...

Страница 28: ...ys in its original package It is very important to pack the Spectro LFP carefully to avoid any dam age during the transport Put foam rubber under the measuring arm This prevents the mea suring arm fro...

Страница 29: ...nds Measuring speed approx 10 min for 1248 fields Repeatability 0 2 Eab 94 on white or 0 005 D up to 1 000 D Inter instrument agree ment typ 1 0 Eab 94 Measuring sensor diffraction grating with diode...

Страница 30: ...rs for special ma terials UVcutfilterreflectionforS3instruments Polarization filter for S3 instruments System requirements Personal Computer running Microsoft Windows XP or higher Apple Intel Mac runn...

Страница 31: ...and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the e...

Страница 32: ...elle presente apparecchiatura giunta a fine vita e organizzata e gestita dal produttore L utente che vorr disfarsi della presente apparecchiatura dovr quindi contattare il produttore e seguire il sist...

Страница 33: ...33 Spectro LFP RT Operating Manual...

Страница 34: ...34 Spectro LFP RT Operating Manual...

Страница 35: ...erating Manual C200M01 C Copyright 2014 by BARBIERI electronic Brixen Italy Information given in this manual is subject to change BARBIERI electronic OHG reserves the right to alter the contents of th...

Страница 36: ...BARBIERI electronic SNC Via Ignaz Seidner 35 39042 BRESSANONE ITALY Tel 39 0472 834 024 Fax 39 0472 833 845 E mail info BARBIERIelectronic com www BARBIERIelectronic com Member of...

Отзывы: