background image

3   

I   

PHscan40 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Temperature Compensation 

In order to obtain accurate measurement results, we recommend that 
enabling the manual temperature compensation before calibration and 
measurement. The tester will automatically calculate the pH slope 
with inputted temperature and shows the temperature compensated 
readings. 
1.  Press and hold the 

Cal

 key to enter the setup menu. 

 

 

 

 
2.  Press the 

Enter

 key until the display shows

 

/

 

 

 

or

 

/

 

 

 

 

 
3.  Press the 

Cal

 key, the tester enters the temperature setting mode. 

4.  Press the 

Cal

 key again to set the temperature value. 

 

 

 

 
5.  Press the 

Enter

 key to save and press the 

Meas

 key to return to 

 

the measurement mode. 

 

 

 

 

 

  During the setting process, press the 

 

Cal 

 key once, the 

 

setting value will increase or decrease by 0.5. Press and hold 

 

the key, the setting value will increase or decrease by 1. 

  To exit the temperature setting without saving

 

changes, press   

 

the 

Meas

 key. 

 

 

pH Calibration 

The PHscan40 tester allows 1 to 3 points calibration, we recommend 
that you perform at least 2 points calibration for accurate pH 
measurement, the tester will automatically recognize and calibrate to 
following standard buffer values. 

 
USA

 

 

standards 

pH 4.01, 7.00, 10.01 

NIST standards 

pH 4.01, 6.86, 9.18 

 

Single point calibration should only be carried out with the pH 7.00 
or 6.86, otherwise calibration will not be accepted. 
 
For better accuracy, we recommend calibrating the tester regularly. 
DO NOT reuse the buffer solutions after calibration, contaminants

 

in

.

 

solution will affect the calibration and eventually the accuracy of

 

the

.

 

measurement. 
 

Single Point Calibration 

Ensure that you have selected 1 point calibration in the setup menu. 
1.1  Press the 

Cal

 key, the tester shows pH 7.00/CAL 1. 

 

 

 

 
1.2  Rinse the electrode with distilled water and place into the pH 
 

7.00 buffer solution, stir the electrode gently to create a   

 

homogeneous solution. 

 

 

 
1.3  Press the 

Enter

 key, the tester begins the calibration. When the 

 

reading has stabilized, the display will show

 

. Calibration 

 

is completed. 

 

 

 

Содержание PHscan40

Страница 1: ...PHscan40 Pocket pH Tester USER MANUAL Bante Instruments Inc ...

Страница 2: ...inimum power requirements the icon automatically disappears MEAS Indicates that the tester is in the measurement mode CAL Indicates that the tester is in the calibration mode SETUP Indicates that the tester is in the setup mode ATC Indicates that the temperature compensation is enabled Prior to Use 1 Take out the pH electrode from the carrying case Insert the BNC connector into the connector socke...

Страница 3: ...ble Disable default Auto Power Off If enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable default Disable Factory Reset If enabled all of the calibration values and current settings will be deleted or reset to the factory defaults the tester must be recalibrated Enable Disable default Setting the Default Option 1 In the measurement mode press and hold the ...

Страница 4: ...bration The PHscan40 tester allows 1 to 3 points calibration we recommend that you perform at least 2 points calibration for accurate pH measurement the tester will automatically recognize and calibrate to following standard buffer values USA standards pH 4 01 7 00 10 01 NIST standards pH 4 01 6 86 9 18 Single point calibration should only be carried out with the pH 7 00 or 6 86 otherwise calibrat...

Страница 5: ...pH 10 01 buffer solution stir the electrode gently Press the Enter key to begin the calibration When the reading has stabilized the display will automatically shows the electrode slope and Calibration is completed During the calibration if the display shows indicating that the measured mV value for the current calibration point deviates by more than 60 mV approximately 1 pH from the theoretical va...

Страница 6: ... deposit by immersing the electrode in warm tap water Rinse the electrode with distilled water and soak in 3M KCl solution Oil or Grease Place the electrode in the detergent or ethanol solution for 15 minutes Rinse the electrode with distilled water and soak in 3M KCl solution Protein 1 Add 1 pepsin to 0 1M HCl solution 2 Place the electrode in above solution for 15 minutes 3 Rinse the electrode w...

Страница 7: ...PHscan40 Range 1 00 to 15 00 pH Resolution 0 01 pH Accuracy 0 01 pH Calibration Point 1 to 3 points pH Buffer Option USA or NIST Automatic Buffer Recognition pH4 01 7 00 10 01or4 01 6 86 9 18 Temperature Compensation 0 to 100 C 32 to 210 F manual Operating Temperature 0 to 50 C 32 to 122 F Storage Temperature 0 to 60 C 32 to 140 F Relative Humidity 80 non condensing IP Rating IP54 Display LCD 21 2...

Страница 8: ...ster is one year from the date of shipment Above warranty does not cover the electrode and pH buffer solutions Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products For mor...