PHscan40 

 

2 Points Calibration 

Make sure that you have selected 2 points calibration in the setup menu. 
2.1  Repeat steps 1.1 through 1.3 above. When the first calibration point is completed, the display will show "----/CAL 2", the tester prompts 
 

you to continue with second point calibration. 

2.2  Rinse the electrode with deionized water and place into the pH

 

4.01 or 10.01 buffer solution. Stir the electrode gently. 

2.3  Press the 

Enter

 key, the tester automatically recognizes the buffer solution and begins calibration. When the reading has stabilized, the 

 

display automatically shows the pH slope (e.g., 99%) and "End". Second point calibration is completed. 

 

 

 

 

 

 

 

3 Points Calibration 

Make sure that you have selected 3 points calibration in the setup menu.   
3.1  Repeat steps 1.1 through 1.3 above. When the first calibration point is completed, the display will show "pH

 

4.01/CAL 2", the tester prompts 

 

you to continue with second point calibration. 

3.2  Rinse the electrode with deionized water and place into the pH

 

4.01 buffer solution. Stir the electrode gently. 

3.3  Press the 

Enter

 key to begin the calibration. When the reading has stabilized, the display will show the pH slope and "pH

 

10.01/CAL 3". 

3.4  Rinse the electrode with deionized water and place into the pH

 

10.01 buffer solution. Stir the electrode gently. When the reading has   

 

stabilized, the display automatically shows the pH slope and "End". Calibration is completed. 

 

   

 

During the calibration, if the display shows

 

 

indicating that the measured mV value for the current calibration point deviates by more 

 

than 60

 

mV (approximately ± 1

 

pH) from the theoretical value of the pH buffer. The calibration will not be accepted. Please check the   

 

electrode and ensure the buffer solutions are fresh and uncontaminated. 

 

To exit the calibration without saving calibrated values, press the 

Meas

 key. 

 

Measurement 

Rinse the electrode with deionized water. Place the electrode into the sample solution and stir gently. Note that the end of the electrode must be 
completely immersed into the solution. Wait for the measurement to stabilize and record the reading. 
 

   

 

During the measurement, never wipe the pH-sensitive glass membrane as this will cause static interference, blot dry with a lint-free tissue 

 

to remove waterdrops on electrode. 

 

If the display shows "----" indicating the measurement exceeds the range, remove the electrode from the sample immediately. 

 

If the

 

 

option is enabled in the setup menu, the tester will automatically lock a measurement endpoint and show

 

 

icon. 

 

Press the 

Meas

 key to resume measuring. 

 