Where: 

 

T

C

  = Temperature coefficient 

 

C

TA

   = Conductivity at temperature A 

 

C

TB

   = Conductivity at temperature B 

 

T

A

    = Temperature A 

 

T

B

    = Temperature B 

 

Calculating the TDS Conversion Factor 

To determine the TDS factor of sample solution use the formula below. 
 

Factor = Actual TDS / Actual Conductivity @25° C

 

 
Where: 
Actual TDS = value from the high purity water and precisely weighed 
 

 

 

   

 

NaCl or KCL reagent 

Actual Conductivity = the meter measured conductivity value 
 
For example: 
Dissolve 64 grams of the potassium chloride (KCl) reagent in 1 liter 
distilled water. If measured conductivity is 100 mS/cm, then TDS factor 
is 0.64. 
 

Optional Accessories 

 
Order Code 

Description 

IE-50LT 

For measuring the pure water 

IE-50MT 

For general purpose applications 

IE-50HT 

For measuring the high conductivity liquids 

ECCS-84 

Conductivity standard solution 84 µ S/cm, 480 ml 

ECCS-1413 

Conductivity standard solution 1413 µ S/cm, 480 ml 

ECCS-1288 

Conductivity standard solution 12.88 mS/cm, 480 ml 

ECCS-1118 

Conductivity standard solution 111.8 mS/cm, 480 ml 

 

Meter Specifications 

 

Model 

BI-650 

Conductivity 

 

Range 

0.01 µ S/cm to 200.0 mS/cm 

Resolution 

0.001, 0.01, 0.1, 1 

Accuracy 

± 1% F.S. 

Calibration Points 

1 to 3 points 

Calibration Solutions 

84 µ S/cm, 1413 µ S/cm, 12.88 mS/cm, 
111.8 mS/cm 

Temperature Compensation 

0 to 100° C (32 to 212° F), automatic 

Temperature Coefficient 

0.0 to 10.0%/° C 

 

 

 

Reference Temperature 

25° C 

Cell Constant 

K = 0.1, 1, 10 

TDS 

 

Range 

0.00 to 10.00 ppt (max. 200 ppt) 

Resolution 

0.01, 0.1, 1 

Accuracy 

± 1% F.S. 

TDS Factor 

0.1 to 1.0 (default 0.5) 

Temperature 

 

Range 

0 to 100° C (32 to 212° F) 

Resolution 

0.1° C (0.1° F) 

Accuracy 

± 1° C (± 1.8° F) 

Calibration Point 

1 point 

Communication 

 

Signal Output 

4 to 20 mA 

Low or High Alarm 

0.02 µ S/cm to 20.0 mS/cm 

Load 

500 ohm 

Communication Interface 

RS485 

Other Specifications 

 

Operating Temperature 

5 to 50° C 41 to 122° F) 

Storage Temperature 

0 to 60° C (32 to 140° F) 

Relative Humidity 

< 80% (non-condensing) 

Display 

LCD, 70 ×  45 mm (2.9 ×  1.7 in.) 

Power Requirements 

DC 24V 

Dimensions 

96 (L) ×  96 (W) ×  75 (H) mm, 
(3.7 ×  3.7 ×  2.9 in.) 

Weight 

350 g (12.3 oz.) 

 
 
 

Содержание BI-650

Страница 1: ...BI 650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc...

Страница 2: ...and operated only in the manner specified in this user manual Only skilled trained or authorized person should carry out installation setup and operation of meter The rear panel of meter has two scre...

Страница 3: ...Display Icon Description Calibration Indicates that the meter is in the calibration mode Setup Indicates that the meter is in the setup mode ATC Indicates that the automatic temperature compensation i...

Страница 4: ...cm default 0 02 S cm Analog Output High Setting range 20 00 mS cm to 0 02 S cm default 20 00 mS cm Factory Reset Reset the meter to factory default settings Note the meter must be recalibrated Enable...

Страница 5: ...K 1 IE 50HT 100 S cm to 200 mS cm K 10 Conductivity Calibration TheBI 650 meter allows 1to 3pointsconductivity calibration Forbetter accuracy we recommend to perform 3 points calibration or select a s...

Страница 6: ...32767 to 32767 hexadecimal PC Command Definition Length of Byte Data ID address 1 0 02 Command 1 0 03 Start address 2 0 0001 Data number 2 0 0002 CRC16 2 0 95F8 Meter Response Definition Length of Byt...

Страница 7: ...Troubleshooting Fault Cause and Corrective Action Electrode dried out Soak the conductivity electrode in tap water for about 10 minutes Measurement exceeded the maximum range Check the electrode and...

Страница 8: ...S cm 480 ml ECCS 1288 Conductivity standard solution 12 88 mS cm 480 ml ECCS 1118 Conductivity standard solution 111 8 mS cm 480 ml Meter Specifications Model BI 650 Conductivity Range 0 01 S cm to 2...

Страница 9: ...eter is one year from the date of shipment Above warranty does not cover the electrode and standard solution Out of warranty products will be repaired on a charged basis The warranty on your meter sha...