18 / 34
23726a GB/2020-07
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
• The
fl
uid in the oscillating tank may not exceed a maximum operating temperature of
100 °C.
• Cover the ultrasonic and rinsing bath when in continuous operation so that not too
much bath liquid evaporates.
4.2
General use
Step 1:
Fill oscillating tank
The oscillating tank is
fi
lled with water and a suitable preparation to reduce the surface
tension, see chapter 7.3.
Fill 1/3 of oscillating tank with water.
Add dosed preparation to the oscillating tank.
Fill carefully up to the
fi
lling level mark, avoid the formation of foam as much as
possible.
Step 2:
Degassing the
fl
uid
Freshly-
fi
lled bath
fl
uid or
fl
uid that has remained in the oscillating tank for a longer
period of time must be degassed prior to use. See also chapter 4.3.1.
Remove basket and other accessories from the oscillating tank.
Place lid on top.
Turn on the ultrasound with the turning knob and sonicate for 30 min for degassing,
see chapter 3.1.1.
Step 3:
Preheat
fl
uid
In ultrasonic baths with integral heating, the
fl
uid may be preheated independently of
the ultrasound. This increases the ultrasound e
ff
ect, especially when removing fats, oils
and polishing paste residue, and shortens the duration of the subsequent ultrasound.
Remove basket and other accessories from the oscillating tank.
Place lid on top.
Using the turning knob, set the desired temperature, see chapter 3.1.2.
For an even warming of
fl
uids, stir the
fl
uids or switch on the ultrasound for a few
minutes now and then, otherwise there will be a retardation of boiling – risk of
scalding!