background image

9

 

/

 

Accessories

34/36

51846-000 en/2022-07

Содержание SONOPULS HD 5020

Страница 1: ...00 en 2022 07 Instructions for use SONOPULS Ultrasonic homogeniser Valid for SONOPULS HD 5020 SONOPULS HD 5050 SONOPULS HD 5100 SONOPULS HD 5200 SONOPULS HD 5400 Dokument 51846 000 Sprache en Version...

Страница 2: ...2 36 51846 000 en 2022 07 2022 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Ger many Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com Certified in accordance with ISO 9001 and ISO 13485...

Страница 3: ...nce of wireless communication 8 3 Design and function 9 3 1 Overview 9 3 2 Symbols and buttons on the touch screen 10 3 3 Function 10 4 Preparation for operation 11 4 1 Set up homogeniser 11 4 2 Mount...

Страница 4: ...5 6 3 Set the language 21 5 6 4 Setting the time and date 21 5 6 5 Set sleep mode 22 5 6 6 Configuring the network 22 5 6 7 Set temperature unit 22 5 7 Troubleshooting a malfunction 23 6 Maintenance 2...

Страница 5: ...ese instructions for use before using the ultrasound bath Pay special attention to chapter 2 Safety If you pass this ultrasound bath on to others please enclose these instruc tions for use Contact you...

Страница 6: ...ours may ignite e g on electrical components resulting in an explosion Do not use combustible solvents in open sonication vessels as the operational safety of the device could be compromised Safe extr...

Страница 7: ...sing a body part to ultrasound can lead not just to skin damage but also to internal tissue damage On fingers the periosteum can become damaged Do not touch the probe the booster horn or the standard...

Страница 8: ...e the probes with care Make sure that the probe does not touch the sonication vessel This can damage the probe and the vessel Foreign bodies on the contact surface between the probe and standard horn...

Страница 9: ...f the device 1 Ultrasound generator 2 Touchscreen 3 Connector for ultrasonic converter 4 LAN connection 5 USB Type B port 6 USB Type A port 7 Mains switch 8 Cold device socket with fuse holder 9 Ultra...

Страница 10: ...e power or amplitude can be selected as the reference variable If the amplitude is selected as the reference variable it is kept constant by the ultrasonic converter regardless of the load provided th...

Страница 11: ...illating system in the housing of the ultrasonic converter twists the ultrasonic converter will be damaged Always use the sickle spanner to secure the standard horn or booster horn Never hold the ultr...

Страница 12: ...the sickle spanner to secure the standard horn or booster horn Never hold the ultrasonic converter by the housing when tightening or loos ening the probe Requirement The ultrasonic converter is discon...

Страница 13: ...asonic converter 4 Place a sickle spanner on the ultrasonic converter Place the ultrasonic con verter on a fixed tabletop and support the sickle spanner on the tabletop 5 Place a sickle spanner on the...

Страница 14: ...he entries with the Confirm Save button Exit the Time Date screen by pressing the Confirm Save button The startup screen appears If the language and time are set the startup screen will appear immedia...

Страница 15: ...nstant The amplitude is given in The absolute value of the amplitude depends on the combination of standard horn or booster horn and probe The power is specified in watts and depends on the respective...

Страница 16: ...hed on and off 5 Confirm your entry with the Confirm Save button 5 2 4 Setting temperature monitoring If you use a temperature sensor you can monitor the temperature of the sonicat ing medium during s...

Страница 17: ...on or end it with the Stop button After the end of sonication a screen with a summary of the sonication appears 4 Tap the Confirm Save button The sonication screen reappears with the last settings 5 4...

Страница 18: ...rtup screen tap Programs The Programs overview screen is shown 2 Tap on a program number that is highlighted in colour The screen with the summary of program parameters appears 3 Tap the Deactivate bu...

Страница 19: ...can still be changed by tapping one of the buttons Amplitude Power Time Pul sation or Temperature Confirm the new settings by tapping the Con firm Save button 3 Optional Tap one of the buttons Amplit...

Страница 20: ...data specified under Institution is stored in the documentation issued 1 On the startup screen tap Settings 2 Tap Documentation 3 Tap Institution 4 Enter the desired data 5 Confirm your entry with the...

Страница 21: ...3 Tap Brightness 4 Set the desired brightness 5 Confirm your entry with the Confirm Save button 5 6 3 Set the language 1 On the startup screen tap Settings 2 Tap on System 3 Tap Language 4 Select the...

Страница 22: ...m Save button 5 6 6 Configuring the network 1 On the startup screen tap Settings 2 Tap on System 3 Tap Interface 4 Tap on Network 5 Activate DHCP if you want the device to automatically obtain the net...

Страница 23: ...place if necessary see 4 3 Mounting removing standard horn or booster horn and 4 2 Mounting removing the probe Erosion on the probe If there is minor erosion repolish the sound emit ting surface of th...

Страница 24: ...ce This leads to a reduction in sonication efficiency The sound emitting surfaces can be carefully polished or milled a few times You can process the surface using a grinder with a fine grinding disc...

Страница 25: ...Before sending the device back to us for inspection repair the device and acces sories must be cleaned pursuant to current laws and regulations and if neces sary must also be disinfected with a surfa...

Страница 26: ...waste if it can no longer be used Do not dispose of the ultrasonic bath with household waste Observe the locally applicable regulations for the disposal of electronic waste The TRISON Base contains a...

Страница 27: ...0 01 99 59 59 Ultrasound operating modes Pulsating continuous Pulsation time 0 5 600 s Regulation Amplitude or power Amplitude setting range 10 100 in 1 steps Power setting power display Watt Energy d...

Страница 28: ...imensions 50 mm 150 mm Degree of protection IP 30 Ultrasonic frequency 20 kHz Weight 0 6 kg Dimensions 50 mm 185 mm Degree of protection IP 30 Ultrasonic frequency 20 kHz Weight 1 0 kg Dimensions 70 m...

Страница 29: ...MC Guideline 2011 65 EU RoHS Directive A declaration of conformity can be requested from the manufacturer by providing the serial number Overvoltage category II Degree of contamination 2 Permissible a...

Страница 30: ...amplitude mSS MS 1 5 1 5 0 1 10 65 MS 2 0 2 0 0 25 20 70 MS 2 5 2 5 0 5 25 72 Probe sound emitting surface mm Sonication volume ml Max amplitude mSS TS 102 2 0 5 20 125 TS 103 3 1 25 118 TS 104 4 5 3...

Страница 31: ...ude mSS TS 103 3 5 90 280 TS 104 4 5 5 100 235 TS 106 6 10 350 220 TS 109 9 10 500 200 TS 113 13 20 900 140 TT 213 13 20 900 105 TS 216 16 25 900 90 TS 219 19 25 900 79 TS 225 25 30 1000 60 Probe soun...

Страница 32: ...ies please refer to the product information appendix to these Instructions for use Stand HG 40 for correct variable positioning of the ultrasonic converter Sound proof box LS 40 reduces noise emission...

Страница 33: ...during mounting disassembling of the probes Sickle spanner HS 40 42 For holding SH 100 G SH 200 G in place during mounting disassembling of the probes Sickle spanner HS 40 42 long 2 pieces Mounting d...

Страница 34: ...9 Accessories 34 36 51846 000 en 2022 07...

Страница 35: ...51846 000 en 2022 07 35 36 Accessories 9...

Страница 36: ...51846 000 en 2022 07 1 2 GmbH Co KG Heinrichstr 3 4 12207 Berlin Germany Phone 49 30 76880 0 Fax 49 30 7734699 info bandelin com www bandelin com...

Отзывы: