probe check
off
scan frequency off
contrast
197
operating mode ampl.
test
US transducer
calibration
probe check
start
scan frequency off
contrast
197
operating mode ampl.
test
US transducer
fine tuning 29910 Hz
START
STOP
off
start
probe check
off
scan frequency off
contrast
197
operating mode ampl.
off
probe check
off
scan frequency start
contrast
197
operating mode ampl.
start
START
STOP
28 / 38
51406c GB/2019-07
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
6
Maintenance and repair
6.1
Maintenance
SONOPULS homogenizers are maintenance-free.
Aside from the process-dependent cavitation erosion on the sound-emitting surface of the probes,
no other homogenizer parts are subject to wear. Worn out or faulty probes can be replaced
following the instructions in chapters 4.3.1.
The devices are calibrated at the time of delivery. A control of the calibration is only required after
repairs, and is only conducted by the manufacturer.
6.2
Functional checks
See also chapter 3.2.3 for description and operation
6.2.1
Testing the ultrasonic transducer (probe check) – Error 011
The test is preferably conducted while the probe is acoustically uncoupled, i.e. the probe should
not be immersed in the sonicating medium.
The HF or ultrasound frequency depends on the probe, and the actual value may vary from the
example presented here. After successful completion, the monitor display switches back to stand-
by mode. Otherwise, an error message appears again.
6.2.2
Conducting a frequency scan (scan frequency) – Error 002/011/012
The conditions for execution are similar to those for conducting checks with the “probe check”
function, see chapter 3.2.3.