The error caused by the thermocouple non-linearity over the  

-50

°

C to 200

°

C range is relatively small, so it is ignored in the 

factory calibration of the TCDA.  If more accurate readings must be 
taken, the following techniques can be used: 
 

1.   If all temperatures are taken near a certain temperature, 

calibrate the TCDA at that particular temperature. 

2.  Implement a look-up table in your software program to 

correct for error. 

3.   Use a piece-wise linear model for your thermocouple output 

to correct for error. 

 
  The terminal blocks on the TCDA and the leads of the 
thermocouple bead are made from dissimilar metals.  This creates a 
second, unwanted thermocouple at the terminal blocks. This 
unwanted thermocouple is called the cold junction.  The temperature 
of this cold junction will affect the results of the actual thermocouple.  
The TCDA has cold junction compensation that automatically 
corrects for small changes in the cold junction temperature.  It is 
important that nothing comes in contact with the terminal blocks or 
the two external diodes near the terminal blocks as this may cause 
large, rapid changes in the cold junction.  The cold junction should 
be kept at a constant temperature, if possible. 
 
 

Different types of thermocouples are made by joining a wide 

variety of metals. The TCDA is designed to amplify K-type 
thermocouple only. Thermocouples have a po) and a 
negative(-) lead.  These two leads must be connected to the correct 
terminals on the TCDA module.  Connections will be covered in the 
next chapter.

Содержание TCDA

Страница 1: ...Designed and Manufactured In Ottawa Illinois USA of domestic and imported parts by B B Electronics Mfg Co Inc 707 Dayton Road P O Box 1040 Ottawa IL 61350 PH 815 433 5100 FAX 815 433 5100 Internet htt...

Страница 2: ...CATIONS 2 CHAPTER 2 THERMOCOUPLE OVERVIEW 3 CHAPTER 3 CONNECTIONS 5 OVERVIEW 5 THERMOCOUPLE CONNECTIONS 5 POWER SUPPLY CONNECTIONS 5 OUTPUT CONNECTIONS 5 CHAPTER 4 CALIBRATION 7 OVERVIEW 7 CALIBRATION...

Страница 3: ...es The amplifier has an output range of 0VDC to 5VDC The standard TCDA can measure temperatures between 50 C and 200 C The module can be modified at the factory so that a different temperature range c...

Страница 4: ...re in your PC 1 Place the disk in drive A 2 Type A and press the ENTER key 3 Type INSTALL and press the ENTER key 4 Follow the instructions given by the program Specifications Size 4 8 x 2 2 x 0 9 Cha...

Страница 5: ...l DC voltage based on the temperature of this junction The TCDA amplifies this small voltage to a usable level for A D converters The output of a thermocouple does not have a linear relationship with...

Страница 6: ...ilar metals This creates a second unwanted thermocouple at the terminal blocks This unwanted thermocouple is called the cold junction The temperature of this cold junction will affect the results of t...

Страница 7: ...2 Make sure that both leads of the thermocouple are connected to the same channel Color coding is used to designate the thermocouple s polarity A K type thermocouple s negative lead is red and the po...

Страница 8: ...0VDC to REF for the SDAXX and SPDA modules provide 5V to REF in the modules AGND to AGND Pin 17 and 18 are internally connected in the TCDA to XXXSDAXX and XXXSPDA A D ch A D ch in the ADIO12 module...

Страница 9: ...ge is relatively small so it is ignored in the factory calibration process If you require more accurate readings you may want to consider using one of the techniques mentioned in Chapter 2 Calibration...

Страница 10: ...e thermocouple in the water as it begins to boil Adjust P1 so that the voltage on DB 25 pin 8 is 3 00VDC adjust P4 so that the voltage on DB 25 pin 9 is 3 00VDC for channel 2 5c If you are using some...

Страница 11: ...2 for channel 2 3 4 Plug power supply in 2 5mm power jack and plug supply into wall outlet 5 Adjust P2 so that there is Voff on pin 6 or 7 of U1 adjust P3 so there is Voff on pin 6 or 7 of U3 for chan...

Страница 12: ...have a modified TCDA module The following values are used for standard TCDA modules 50 C Temperature 200 C 0VDC Vout 5VDC Vout is measured on DB 25 pin 8 pin 9 for channel 2 resolution 20mV C 0 02V C...

Страница 13: ...the example files on your hard drive Place the disk in drive A Type A and press the ENTER key Type INSTALL and press the ENTER key Follow the instructions given by the program After the software inst...

Отзывы: