436 M1 01 C
+ mA ouput / measurement
- mA ouput / measurement
- mA ouput / temperature
+ mA ouput / temperature
+ 24 V
To switch the regulation on stand-by
- 24 V
- 24 V
+ 24 V
Power supply = 20 mA
Pt 100 Ω
Shield
Y
Z
Connection to an extension terminal (blind BAMOPHOX)
B
A
S4 contact, NO, alarm on temperature or over range
S2 threshold, NO contact – Measurement / Temperature
S1 threshold NO contact – Measurement / Temperature
S3 threshold NO contact – Measurement / Temperature
NOT USED
Ground
N
Main power supply (N= Neutral, L= Line)
L
NC
- mA
Turbidity transmitter
+ mA
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
Screwing terminals
P +
P -
+
4-20 mA
-
TURBICUBE
or
GAB
Wiring from wall or panel mounting BAMOPHOX to an Extension terminal BAMOPHOX
- Maximum length cable:
500 m
- Wire specifications:
Mains cable or 4 wires
shielded cable
≥ 0,25 mm
2
cross section
Main unit
Y
Z
B
A
13
14
15
16
Extension terminal
Y
Z
B
A
13
14
15
16