background image

BG 800 300 BE / A     (0105)     TPG 300

23

Interface

Baud rate

3     (300)
1   (1200)
2   (2400)
4   (4800)
9   (9600)

Parameter storage

Parameter program selection

u    (user)

     Code

     00 0 ... 99 19

H   (hotstart)

     Code

     00 0 ... 99 19

d    (default)

Storage (save)

Time for consideration

Test programs

Test program selection

dI     (display)
rA    (RAM)
EP   (EPROM)
EE   (EEPROM)
Ad   (A/D converter)

     Channel selection

     A0 .. A7

Io     (keys)
rS    (interface)
Pn   (program number)

Test

t      (test start)

 

 

 

 

 

Содержание TPG 300

Страница 1: ...Operating manual BG 800 300 BE A 0105 1 Total pressure gauge and controller TPG 300 Configuration Installation Operation Technical data Program pg48...

Страница 2: ...plate This manual is based on firmware version BG 509 780 F Proceed according to paragraph 4 7 10 to check that your unit is equipped with this or a higher firmware version F or higher Enter the firmw...

Страница 3: ...4 1 5 Alignment 12 4 2 Front panel 13 4 3 Operating modes 14 4 4 Overview of key entries 15 4 5 sensor mode 16 4 5 1 Switching the measuring circuit on off 16 4 5 2 Measurement range violation 17 4 5...

Страница 4: ...1 Installation and connection diagram 44 8 2 Data transmission 44 8 2 1 Definitions 44 8 2 2 Flow control 45 8 2 3 Communication protocol 45 8 3 Mnemonics 47 8 3 1 Measured values 48 8 3 2 Switching...

Страница 5: ...NOTE Special information on cost effective use Specialists This work may only be carried out by persons with suit able technical training and the necessary experience Press this key briefly Press the...

Страница 6: ...esponsibility nor warranty if the user or third parties utilize the product not according to the defined use make any kind of changes modifications alterations etc to the product use the product with...

Страница 7: ...ation of the plug in boards 3 A list of all plug in boards suited for the TPG 300 can be found in section 5 3 For detailed information on the plug in boards 3 Interface plug in boards Relay and RS 232...

Страница 8: ...asurement plug in boards Dual Pirani e g PI 300D Cold cathode e g PE 300C9 PE 300T10 PE 300DC10 Pirani cold cathode combined e g CP 300C9 TPR 010 TPR 017 TPR 018 IKR 050 IKR 060 IKR 070 TPR 010 TPR 01...

Страница 9: ...1 Mains power connection You can connect the TPG 300 without voltage adaptation to any conventional supply system section 5 1 The power connector may only be plugged into a socket with a protective gr...

Страница 10: ...in board Units with PI 300D and PE 300 measurement plug in boards The cold cathode measuring circuit is controlled by the Pirani measuring circuit TPR 2 The controlling Pirani gauge and the controlle...

Страница 11: ...ld rack frames constructed according to the Balzers standard Older units Balzers standard are to be installed in the same rack frame as the TPG 300 DIN a special adapter must be used section 9 6 3 5 U...

Страница 12: ...the measurement value of the first measuring circuit in opera tion is displayed 4 1 2 Gas type dependence The measured pressure depends on the gas type present It is referenced to nitrogen N2 For othe...

Страница 13: ...group TPG 300 1 2 3 4 A B mbar A1 PE Err 0 10 Unit of measurement Selected sensor display Error signal Select sensor for display of measured data Status of switching function lights if status ON Chan...

Страница 14: ...sor Pressure measurement Selection of the measuring circuit Switching gauges on off set point Display of the switching function parameters set up Display of the unit parameters Modification of the uni...

Страница 15: ...e Change to set up mode Select next switching function in set point Change to sensor mode Change to set up mode Select next group in set up Select next function in set up if existing Modify the select...

Страница 16: ...p and funct Manual on off switching has priority over the automatic control Measured value 1 2 3 4 A B mbar A1 0 10 Plug in board identification 1 2 3 4 A B A1 0 10 Switch on cold cathode gauges at pr...

Страница 17: ...s play changes over automatically Overrange 1 2 3 4 A B mbar B1 PE 0 10 Underrange 1 2 3 4 A B mbar B1 PE 0 10 If the upper measuring range limit is exceeded the cold cathode gauge can become contamin...

Страница 18: ...threshold Measured value or e g 1 2 3 4 A B B1 0 10 Automatic control Au cold cathode measuring circuit waits for the fulfillment of the power on condition by the Pirani measuring circuit 4 5 4 Self...

Страница 19: ...d cathode measuring circuit 1 10 10 mbar automatic operation Cold cathode measuring circuit 10 11 mbar automatic operation PE 9 PE 10 PE 11 Cold cathode measuring circuit 5 10 9 mbar Cold cathode meas...

Страница 20: ...back to sensor mode measuring mode Assignment of switching function to meas uring circuit With set point you can read cyclically the threshold values and assignments of the switching functions 1 2 3...

Страница 21: ...groups group and functions funct Select the parameter to be displayed in the set up mode Push the group or funct key Modify the parameter by pushing the step key 4 7 2 Parameter overview Comments to t...

Страница 22: ...er upper Threshold 1st digit mantissa 1 9 Threshold 2nd digit mantissa 0 9 Threshold exponent 11 3 Measuring circuit assignment A1 A2 B1 B2 none PE measuring circuit underrange control Control 0 disab...

Страница 23: ...age Parameter program selection u user Code 00 0 99 19 H hotstart Code 00 0 99 19 d default Storage save Time for consideration Test programs Test program selection dI display rA RAM EP EPROM EE EEPRO...

Страница 24: ...r Switching function 4 1 0 10 11 mbar Switching function A 6 0 10 3 mbar Switching function B 6 0 10 3 mbar Upper threshold Switching function 1 9 0 10 11 mbar Switching function 2 9 0 10 11 mbar Swit...

Страница 25: ...ium Measuring circuit B1 2 medium Measuring circuit B2 2 medium Baud rate 9 9600 Immediate start up Measuring circuit A1 no Measuring circuit A2 no Measuring circuit B1 no Measuring circuit B2 no Code...

Страница 26: ...rface plug in board relay 4 Cold cathode measuring circuit s in slot A Cold cathode measuring circuit s in slot B b Threshold selection function Defining an upper and a lower threshold defines a hyste...

Страница 27: ...e depends on the plug in boards installed Available measuring circuits are indicated by an LED The upper and lower thresholds of switching functions 1 4 cannot be assigned to different measuring circu...

Страница 28: ...error the switching function changes to off The switching function does not change to on until the measured value has remained within the measurement range of the cold cathode measuring circuit for at...

Страница 29: ...talled b Filter time constant function Three filter time constants are available In the case of signal fluctuations a faster filter can cause fluttering of switching functions 1 2 3 4 A B A1 0 10 Disp...

Страница 30: ...default factory set parameters SAve de faults By activating the immediate start up a measuring circuit can be automatically re enabled after a power failure This is particularly useful in the case of...

Страница 31: ...do not wish to modify the lock then push funct three times Pushing step until the bar graph is completely dark causes the parameters to be stored When the storage process is finished the bar graph lig...

Страница 32: ...rter Channel 0 Channel 1 Channel 2 Channel 3 Channel 4 Channel 5 Channel 6 Channel 7 Keys RS 232 C interface Program number The display test lights first all LEDs together and then in dividually The R...

Страница 33: ...ushing funct or group All the other tests run through once When they are finished a line appears in the exponent display You can stop the dI by pushing step and start it again as often as you like c P...

Страница 34: ...C decreasing to 50 at 40 C Use indoors only Height up to 2000 m Protection IP 40 Safety EN 61010 93 IEC 1010 90 A1 92 Class I Pollution degree 2 EMC Emission EN 50081 1 92 Immunity EN 50082 2 95 5 3 B...

Страница 35: ...10N CP 300T10 CP 300T10N CP 300T10L CP 300T10LN CP 300T11 CP 300T11N CP 300T11L CP 300T11LN Compatible interface plug in boards IF 300A IF 300B IF 300C 5 4 Dimensions and weight Dimensions sensor set...

Страница 36: ...ing Err LED 1 2 3 4 A B mbar A2 Err 0 10 Display Possible cause Correction dt 3 Watch Dog timer overflow due to strong external influ ence electromagnetic funct If this error occurs frequently replace...

Страница 37: ...rogram Interface defective Replace interface plug in board 2 SE 4 TPR gauge not connected Connect gauge Break in TPR cable Replace cable TRR gauge defective Replace gauge So 3 Stack overflow funct 2 P...

Страница 38: ...ized in the event of a fault in the assigned measuring circuit a fatal error Contact 5 Error status is de energized in the event of a fault in a measuring circuit a fatal error Additional information...

Страница 39: ...suring circuit Clean the gauge 4 6 Replace the gauge Pirani measurement circuit cannot be cali brated Combination measurement plug in board gauge cable gauge is not compatible Select correct combinati...

Страница 40: ...s approx 1 cm out of the slots A and B 2 2 Select the code in set up mode 3 Read out the code 4 Reinstall the measurement plug in boards 2 2 Please read the instructions in 3 2 2 before performing any...

Страница 41: ...e back panel Pull out the back panel together with the plug in boards just far enough to make the fuses accessible Replace the fuses Make sure to replace fuses with fuses of the recom mended type and...

Страница 42: ...of the work described below For cleaning the outside of the unit a slightly humid cloth will usually do Do not use under any circumstances an aggressive or scouring leaning agent No water must get in...

Страница 43: ...the electronics of the TPG 300 from static electricity Procedure 1 Store default values 2 30 2 Switch the unit off and wait one minute 3 Detach all cables from the unit mains cable last 4 Unscrew the...

Страница 44: ...her direction 8 2 1 Definitions The following abbreviations and symbols are used Symbol Meaning HOST Computer or terminal Non mandatory elements ASCII American Standard Code for Information Interchang...

Страница 45: ...le value ranges are also specified there Send protocol HOST TPG 300 Explanation Mnemonics and parameters CR LF Receives message with end of message ACK CR LF Positive acknowledgment of a received mess...

Страница 46: ...RROR word Errors can be decoded after the ERROR word has been read Error recognition protocol HOST TPG 300 Explanation Mnemonics and parameters CR LF Receives message with end of message Transmission...

Страница 47: ...SP2 SP3 SP4 SPA SPB SPS Set point 1 Set point 2 Set point 3 Set point 4 Set point A Set point B Set point status Switching function 1 Switching function 2 Switching function 3 Switching function 4 Sw...

Страница 48: ...ing circuit A2 2 automatic Measuring circuit A1 3 on Receive ACK CR Send ENQ Receive x x x x CR LF Status measuring circuit B2 Status measuring circuit B1 Status measuring circuit A2 Status measuring...

Страница 49: ...lue 1 0E 11 to 1 4E 3 Status x 0 Measurement data okay 1 Underrange 2 Overrange 3 Measuring circuit error 4 Measuring circuit switched off 5 No hardware PE underrange control Send PUC x CR LF PE under...

Страница 50: ...2 Measuring circuit A2 3 Measuring circuit B1 4 Measuring circuit B1 5 Measuring circuit A1 6 Measuring circuit A2 7 Measuring circuit B1 8 Measuring circuit B1 upper threshold 9 9E 3 1 0E 11 lower th...

Страница 51: ...function assignment upper threshold lower threshold Switching function status Send SPS CR LF Set point status Receive ACK CR Send ENQ Receive x x x x x x CR LF Switching function B x 0 off Switching f...

Страница 52: ...00 300 BE A 0105 TPG 300 52 8 3 3 Display Unit of measurement pressure Send UNI x CR LF Unit of measurement Unit of measure x 1 mbar 2 Torr 3 Pa Receive ACK CR Send ENQ Receive x CR LF Unit of measure...

Страница 53: ...lter time constant Measuring circuit B2 x 1 fast Measuring circuit B1 2 medium Measuring circuit A2 3 slow Measuring circuit A1 Receive ACK CR Send ENQ Receive x x x x CR LF Filter measuring circuit B...

Страница 54: ...2400 baud 4 4800 baud 9 9600 baud Receive ACK CR LF Send ENQ Receive x CR LF Baud rate 8 3 6 Storing Send SAV x CR LF Save parameters Set of parameters x 0 Default parameters 1 User parameters 2 User...

Страница 55: ...without entering a code 1 1998 TPG can be operated when a corresponding code is entered 1999 TPG can be operated only when this particular code is entered code cannot be modified Receive ACK CR LF Sen...

Страница 56: ...dentification Receive ACK CR Send ENQ Receive x x x CR LF Identification slot C Identification slot B Identification slot A x i e NO P no plug in board PI300 DL Test programs for Balzers Service 8 3 8...

Страница 57: ...S E S E TID CR LF ACK CR LF ENQ PI 300DL PE 300 10 IF 300 CR LF Request for TPG identification Positive acknowledgment Inquiry Plug in board types S E S E SEN CR LF ACK CR LF ENQ 3 3 1 0 CR LF Reques...

Страница 58: ...Request check of sensor status Positive acknowledgment Inquiry Sensor status report S E SAV 1 CR LF ACK CR LF Store modified set of parameters Positive acknowledgment S E S E S E PA2 CR LF ACK CR LF...

Страница 59: ...CS DS CD AS 1 21 REM Eroeffnet COM1 mit 9600 bps keine Paritaet und acht Daten Bits 22 REM CTS DSR und CD werden nicht geprueft 23 REM 30 ACK CHR 6 ENQ CHR 5 LF CHR 10 100 LINE INPUT Mnemonics m 101...

Страница 60: ...ATAL ERROR E E 1000 240 IF e 99 THEN PRINT NO HARDWARE E E 100 250 IF e 9 THEN PRINT PARAMETER ERROR E E 10 260 IF e THEN PRINT SYNTAX ERROR 270 PRINT 280 GOTO 100 300 END 10 REM 11 REM AVMTPG BAS Abf...

Страница 61: ...tus 320 GOSUB 1000 REM Kommunikationsprotokoll 330 PRINT 1 ENQ REM Aufforderung zur Datenuebertragung 340 INPUT 1 r1 r2 r3 r4 ra rb REM Lesen des Status 350 PRINT USING R1 R2 R3 R4 RA RB r1 r2 r3 r4 r...

Страница 62: ...in boards For ordering numbers of the compatible measurement and interface plug in boards c f section 5 3 3 9 3 Mains cables Ordering number Mains cable Swiss standard Type 432 2 5 m Mains cable Schu...

Страница 63: ...ion in a rack Ordering number 19 rack frame kit DIN 41 494 3HE 84TE for new submodules for combination with older units Balzers standard without front profile noses 2 pairs of fixing bars tapped at 5...

Страница 64: ...e Err EE EP Test program error message Err Test program error message Err FI Filter Id IF Io Error message Error message Test program error message Err or Overrange PE PI Pn Cold cathode measuring cir...

Страница 65: ...g instructions TPR 017 Pirani gauge head BG 800 317 BE 2nd edition 6 1991 Balzers AG 9496 Balzers F rstentum Liechtenstein 6 Operating instructions TPR 018 Pirani gauge head BG 800 318 BE 2nd edition...

Страница 66: ...ter assignment 29 filter time constant 29 overview 21 parameter set 30 program version 33 save 30 test 33 test program selection 32 G Gas type dependence 12 Gauge cables 63 switching off 18 Gauges 63...

Страница 67: ...anging the 14 Operating modes 14 Overview default parameters 24 functions 21 groups 21 key entries 15 operating modes 14 parameters 21 system 7 P Parameters default 24 30 storage 30 Plug in boards 62...

Страница 68: ...vormals Balzers Aktiengesellschaft LI 9496 Balzers Liechtenstein Tel 423 388 3111 Fax 423 388 3700 Original German BG 800 300 BD A reach liechtenstein inficon com bg800300be a www inficon com...

Отзывы: