D30R Users Manual
129
Do Not Touch Test Leads while test is in Progress!
71-023 Rev C 5/8/2007
nose pliers. Baker specifies the following fuse rating: 5x20mm 250V 2.5A
Slow-Blow.
12.
The screen shows a wavy appearance when the unit is at or near maximum
output surge test voltage.
The AC power source needs to supply at least 200 watts with minimal sag. If
the AC supply is extremely weak, the CRT may show some wavy appearing
lines. Use a power supply rated at least to 500 watts for reliable, trouble-free
performance.
Service: What to do First
?
Because history has shown that several simple solutions which do not require
return of an unit may arise, please perform the following checks.
Open Condition Display
Note the figures below. Is the surge waveform like this?
If yes, the unit may
have at least one
broken test lead
causing an Open
condition. In most
cases, the test lead in
common with the two
TLS positions that
produces these types
of wave patterns is the
lead which is broken.
Verify this by pulling
on the book/clip
assembly of the lead. A broken test lead will stretch. If it does not repeat this
procedure at one foot intervals for the length of the lead. If the leads of the
tester are good, check the connections and continuity of the test winding.
Fig B-1: Open Condition Display
Содержание D30R
Страница 2: ...D30R Users Manual 2 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 4: ...D30R Users Manual 4 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 10: ...D30R Users Manual 10 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 16: ...D30R Users Manual 16 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 24: ...D30R Users Manual 24 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 32: ...D30R Users Manual 32 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 34: ...D30R Users Manual 34 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 40: ...D30R Users Manual 40 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 46: ...D30R Users Manual 46 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 62: ...D30R Users Manual 62 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 68: ...D30R Users Manual 68 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 70: ...D30R Users Manual 70 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 80: ...D30R Users Manual 80 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 82: ...D30R Users Manual 82 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 106: ...D30R Users Manual 106 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 108: ...D30R Users Manual 108 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 118: ...D30R Users Manual 118 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 120: ...D30R Users Manual 120 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 124: ...D30R Users Manual 124 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 134: ...D30R Users Manual 134 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 136: ...D30R Users Manual 136 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Страница 139: ...D30R Users Manual 139 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007 INDEX...
Страница 140: ...D30R Users Manual 140 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...