SI-46 Version 2.0 

Page 3 of 34 

SCI-tive Service Procedure 

1.

 

Overview 

1.1

 

Safety Instructions 

Baker and/or Ruskinn do not take any responsibility for damages caused by using the equipment for 
other purposes than described in this user manual. 

 

The mains appliance coupler and plug are the AC mains supply isolation device and must be easily 
accessible when installed. 

 

In case of emergency disconnect the SCI-tive from the AC Mains Outlet. 

 

Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved. 

 

All installation work and adjustments to the unit must be carried out by qualified personnel. Work 
performed by persons with insufficient technical knowledge may adversely affect the performance 
of the unit or cause physical injury or damage to the equipment. 

 

All servicing and repairs must be carried out by a qualified customer service engineer. Only genuine 
spare parts must be used. 

 

In case of damage to the SCI-tive disconnect the workstation from the mains outlet and contact 
your local distributor. 

 

All covers and lids must only to be removed by a qualified service engineer. 

 

Nothing should be placed on the top of the workstation. 

 

A power cord supplied with the workstation and should be used to connect to the mains outlet. If 
a replacement is required it must be adequately rated for the application. 

 

All cables and pipes should be routed to ensure that they do not pose a trip hazard. 

 

Mains supply Voltage fluctuation must not exceed ±10% of the nominal Voltage. 

 

Gas  regulators  must  be  used  for  each  gas  supply.  A  2  stage  regulator  is  recommended  with  a 
maximum supply pressure of 4 Bar. Over pressure could damage the workstation. 

 

Only the gases specified in this user manual may be used. 

 

All gas bottles must be adequately secured before connection to the workstation. 

 

The maximum power rating of the internal sockets must not be exceeded. 

 

The end user is responsible for all materials and equipment places inside the workstation. 

 

Before connecting any equipment to the internal mains sockets ensure that the equipment has or 
is connected to a protective earth. If in doubt please contact the equipment manufacturer.  

 

The workstation must not be operated at an ambient temperature over 30°C. 

 

The cooling fan covers and cooling vents must not be covered or blocked. 

 

The weight limit for the workstation internal floor is 30kg and must be evenly distributed. 

 

The weight limit for the interlock tray is 5kg and must be evenly distributed. 

 

There should be no naked flames close to the workstation. 

 

The use of radioactive materials if strictly prohibited. 

 

CAUTION: Asphyxiation Risk 

The  SCI-tive  uses  Nitrogen  (N

2

)  and  Carbon  Dioxide  (CO

2

)  as  part  of  normal  use  with  the  volume 

released  externally  is  inconsequential.  In  the  event  of  a  leak  or  malfunction  this  gas  release  may 
become excessive. DO NOT OPERATE this unit in a SMALL ENCLOSURE such as a small room or walk-
in  closet.  An  accidental  release  of  Nitrogen  or  Carbon  Dioxide  could  create  an  asphyxiating 
atmosphere in a small space. 

 

If the equipment is not use in a manor specified by the manufacture, the protection provided by the 
equipment may be impaired. 

 

Failure to adhere to these safety instructions could cause serious injury and will invalidate the 
workstation warranty. Ruskinn technology limited accepts no responsibility for any accident, injury 
or loss caused by unsafe operation of the workstation. 

Содержание Sci-tive SI-046

Страница 1: ......

Страница 2: ...TRAY 17 3 2 REMOVING THE CEILING PANEL 18 3 3 REMOVING THE REAR WALL 18 3 4 REPLACING THE CIRCULATION FANS 19 3 5 REPLACING THE INTERLOCK DOOR CLAMPS 20 3 6 REFITTING THE REAR WALL 21 3 7 REPLACING TH...

Страница 3: ...maximum supply pressure of 4 Bar Over pressure could damage the workstation Only the gases specified in this user manual may be used All gas bottles must be adequately secured before connection to the...

Страница 4: ...tronic Equipment WEEE Directive This product must not be treated as household waste Instead it shall be handed over to an appropriate collection point for the recycling of electrical and electronic eq...

Страница 5: ...ea is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense 1 2 4 Precautions The SCI tive has been tested and approved to EN61010...

Страница 6: ...rs No end user serviceable parts behind covers Please refer to this manual in all cases where this symbol appears in order to find out the nature of the Potential Hazard and actions to be taken in ord...

Страница 7: ...SI 46 Version 2 0 Page 7 of 34 SCI tive Service Procedure 2 1 Year Service Procedure 2 1 Removing the Detox Sachets 2 2 Replacing the Oxygen Sensor o Figure 2 Humidity End Panel...

Страница 8: ...SI 46 Version 2 0 Page 8 of 34 SCI tive Service Procedure Figure 3 Oxygen Sensor...

Страница 9: ...SI 46 Version 2 0 Page 9 of 34 SCI tive Service Procedure 2 3 Cleaning the Water and Oil Tanks Figure 5 Water tank drain tube Figure 4 Inside panel of water tank 1...

Страница 10: ...SI 46 Version 2 0 Page 10 of 34 SCI tive Service Procedure Figure 6 Water tank HEPA filter tube connections Figure 7 By passing water tank HEPA filter 1 2 3...

Страница 11: ...SI 46 Version 2 0 Page 11 of 34 SCI tive Service Procedure Figure 8 Secondary Pressure Relief oil tank 1 2...

Страница 12: ...SI 46 Version 2 0 Page 12 of 34 SCI tive Service Procedure Figure 9 Engineering Mode Software Screen Figure 10 Engineering 2 Software screen...

Страница 13: ...SI 46 Version 2 0 Page 13 of 34 SCI tive Service Procedure Figure 11 Humidity System Settings...

Страница 14: ...SI 46 Version 2 0 Page 14 of 34 SCI tive Service Procedure 2 4 Replacing the Humidity Sensor Figure 12 Humidity Chip Removable Cover Plate...

Страница 15: ...SI 46 Version 2 0 Page 15 of 34 SCI tive Service Procedure 2 5 Testing procedure o o o...

Страница 16: ...SI 46 Version 2 0 Page 16 of 34 SCI tive Service Procedure Figure 13 Pressure Integrity Test Screen...

Страница 17: ...SI 46 Version 2 0 Page 17 of 34 SCI tive Service Procedure 3 2 Year Service 3 1 Removing the Floor Tray...

Страница 18: ...SI 46 Version 2 0 Page 18 of 34 SCI tive Service Procedure 3 2 Removing the Ceiling Panel o Figure 14 Ceiling Panel 3 3 Removing the Rear Wall Figure 15 Rear Wall...

Страница 19: ...SI 46 Version 2 0 Page 19 of 34 SCI tive Service Procedure 3 4 Replacing the Circulation Fans Figure 16 Circulation Fans 1...

Страница 20: ...ion 2 0 Page 20 of 34 SCI tive Service Procedure 3 5 Replacing the Interlock Door Clamps Figure 17 Inner Door Switch Moulded Cover Figure 18 Inner Door Clamp Brackets Figure 19 Inner Door Tube Retaini...

Страница 21: ...SI 46 Version 2 0 Page 21 of 34 SCI tive Service Procedure Figure 20 Inner Door Clamps 3 6 Refitting the Rear Wall...

Страница 22: ...SI 46 Version 2 0 Page 22 of 34 SCI tive Service Procedure 3 7 Replacing the Oxygen Sensor o Figure 21 Humidity End Panel Figure 22 Oxygen Sensor...

Страница 23: ...SI 46 Version 2 0 Page 23 of 34 SCI tive Service Procedure 3 8 Replacing the Condenser Fan Figure 23 Condensate Fan...

Страница 24: ...24 of 34 SCI tive Service Procedure 3 9 Replacing the Gas Sample Pump o Figure 24 Gas Sample Pump 3 10 Replacing the Gas Sample Condensate Fan Figure 25 Gas Sample Condensate Fan Figure 26 Gas Sample...

Страница 25: ...Replacing the Gas Sample Heatsink Fan Figure 27 Gas Sample Heatsink Fan Location Figure 28 Gas Sample Heatsink Fan Fixings 3 12 Replacing the Interlock End Panel Air Circulation Fan Figure 29 Interlo...

Страница 26: ...n 2 0 Page 26 of 34 SCI tive Service Procedure 3 13 Cleaning the Water and Oil Tanks Figure 32 Water tank drain tube Figure 33 Water tank HEPA filter tube connections Figure 31 Inside panel of water t...

Страница 27: ...SI 46 Version 2 0 Page 27 of 34 SCI tive Service Procedure Figure 34 Bypassing water tank HEPA filter...

Страница 28: ...SI 46 Version 2 0 Page 28 of 34 SCI tive Service Procedure Figure 35 Secondary Pressure Relief oil tank 1 2...

Страница 29: ...SI 46 Version 2 0 Page 29 of 34 SCI tive Service Procedure Figure 36 Engineering Mode Software Screen Figure 37 Engineering 2 Software screen...

Страница 30: ...SI 46 Version 2 0 Page 30 of 34 SCI tive Service Procedure Figure 38 Humidity System Settings...

Страница 31: ...SI 46 Version 2 0 Page 31 of 34 SCI tive Service Procedure 3 14 Replacing the Humidity Sensor Figure 39 Humidity Chip Removable Cover Plate...

Страница 32: ...SI 46 Version 2 0 Page 32 of 34 SCI tive Service Procedure 3 15 Testing procedure o o o...

Страница 33: ...SI 46 Version 2 0 Page 33 of 34 SCI tive Service Procedure Figure 40 Pressure Integrity Test 4 Reference Documents...

Страница 34: ...SI 46 Version 2 0 Page 34 of 34 SCI tive Service Procedure...

Отзывы: