SI-46 Version 2.0
Page 3 of 34
SCI-tive Service Procedure
1.
Overview
1.1
Safety Instructions
Baker and/or Ruskinn do not take any responsibility for damages caused by using the equipment for
other purposes than described in this user manual.
The mains appliance coupler and plug are the AC mains supply isolation device and must be easily
accessible when installed.
In case of emergency disconnect the SCI-tive from the AC Mains Outlet.
Ensure that the connecting cable is not squeezed or bent when the unit is being installed or moved.
All installation work and adjustments to the unit must be carried out by qualified personnel. Work
performed by persons with insufficient technical knowledge may adversely affect the performance
of the unit or cause physical injury or damage to the equipment.
All servicing and repairs must be carried out by a qualified customer service engineer. Only genuine
spare parts must be used.
In case of damage to the SCI-tive disconnect the workstation from the mains outlet and contact
your local distributor.
All covers and lids must only to be removed by a qualified service engineer.
Nothing should be placed on the top of the workstation.
A power cord supplied with the workstation and should be used to connect to the mains outlet. If
a replacement is required it must be adequately rated for the application.
All cables and pipes should be routed to ensure that they do not pose a trip hazard.
Mains supply Voltage fluctuation must not exceed ±10% of the nominal Voltage.
Gas regulators must be used for each gas supply. A 2 stage regulator is recommended with a
maximum supply pressure of 4 Bar. Over pressure could damage the workstation.
Only the gases specified in this user manual may be used.
All gas bottles must be adequately secured before connection to the workstation.
The maximum power rating of the internal sockets must not be exceeded.
The end user is responsible for all materials and equipment places inside the workstation.
Before connecting any equipment to the internal mains sockets ensure that the equipment has or
is connected to a protective earth. If in doubt please contact the equipment manufacturer.
The workstation must not be operated at an ambient temperature over 30°C.
The cooling fan covers and cooling vents must not be covered or blocked.
The weight limit for the workstation internal floor is 30kg and must be evenly distributed.
The weight limit for the interlock tray is 5kg and must be evenly distributed.
There should be no naked flames close to the workstation.
The use of radioactive materials if strictly prohibited.
CAUTION: Asphyxiation Risk
The SCI-tive uses Nitrogen (N
2
) and Carbon Dioxide (CO
2
) as part of normal use with the volume
released externally is inconsequential. In the event of a leak or malfunction this gas release may
become excessive. DO NOT OPERATE this unit in a SMALL ENCLOSURE such as a small room or walk-
in closet. An accidental release of Nitrogen or Carbon Dioxide could create an asphyxiating
atmosphere in a small space.
If the equipment is not use in a manor specified by the manufacture, the protection provided by the
equipment may be impaired.
Failure to adhere to these safety instructions could cause serious injury and will invalidate the
workstation warranty. Ruskinn technology limited accepts no responsibility for any accident, injury
or loss caused by unsafe operation of the workstation.
Содержание Sci-tive SI-046
Страница 1: ......
Страница 8: ...SI 46 Version 2 0 Page 8 of 34 SCI tive Service Procedure Figure 3 Oxygen Sensor...
Страница 11: ...SI 46 Version 2 0 Page 11 of 34 SCI tive Service Procedure Figure 8 Secondary Pressure Relief oil tank 1 2...
Страница 13: ...SI 46 Version 2 0 Page 13 of 34 SCI tive Service Procedure Figure 11 Humidity System Settings...
Страница 15: ...SI 46 Version 2 0 Page 15 of 34 SCI tive Service Procedure 2 5 Testing procedure o o o...
Страница 16: ...SI 46 Version 2 0 Page 16 of 34 SCI tive Service Procedure Figure 13 Pressure Integrity Test Screen...
Страница 17: ...SI 46 Version 2 0 Page 17 of 34 SCI tive Service Procedure 3 2 Year Service 3 1 Removing the Floor Tray...
Страница 27: ...SI 46 Version 2 0 Page 27 of 34 SCI tive Service Procedure Figure 34 Bypassing water tank HEPA filter...
Страница 28: ...SI 46 Version 2 0 Page 28 of 34 SCI tive Service Procedure Figure 35 Secondary Pressure Relief oil tank 1 2...
Страница 30: ...SI 46 Version 2 0 Page 30 of 34 SCI tive Service Procedure Figure 38 Humidity System Settings...
Страница 32: ...SI 46 Version 2 0 Page 32 of 34 SCI tive Service Procedure 3 15 Testing procedure o o o...
Страница 34: ...SI 46 Version 2 0 Page 34 of 34 SCI tive Service Procedure...