Baker Instrument Company D12R Скачать руководство пользователя страница 1

 

2/24/2010 | 71-018 V11 EN  
D12R, D6R, D3R Users Manual 

 

Users Manual 

Digital Winding Tester 

D12R, D6R, 3R 

 

Baker Instrument Company, an SKF Group Company, 

4812 McMurry Ave. Suite 100 
Fort Collins, CO 80525 
(970) 282-1200 
(970) 282-1010 (FAX) 
800-752-8272 (USA Only) 
 
Note: This equipment has been tested and found to comply with the limits for a Class A digital 
device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable 
protection against harmful interference with the equipment is operated in its installation. This 
equipment generates uses and can radiate radio frequency energy and, if not installed and used in 
accordance with the product manual, may cause harmful interference to radio communications. If 
this equipment does cause harmful interference, the user will be required to correct the 
interference. 
 
Due to the phenomena being observed and the material properties being measured, this 
equipment does radiate radio frequency energy while in the active test mode. Care should be taken 
to insure this radio frequency energy causes no harm to individuals or other nearby equipment. 

 

 
Information furnished in this manual by Baker Instrument Company, an SKF Group Company,, is 
believed to be accurate and reliable. However, Baker Instrument Company, an SKF Group 
Company, assumes no responsibility for the use of such information or for any infringements of 
patents or other rights of third parties that may result from its use.  No license is granted by 
implication or otherwise under any patent rights of Baker Instrument Company, an SKF Group 
Company. 

 
Warning:

  

Baker Instrument Company, an SKF Group Company, assumes no liability for damages consequent 
to the use of this product. No part of this document may be reproduced in part or in full by any 
means such as photocopying, photographs, electronic recording, videotaping, facsimile, etc., 
without written permission from Baker Instrument Company, an SKF Group Company, Fort Collins, 
Colorado. 

 

Intended use of instrument 

The D12R, D6R and D3R, is offered by Baker Instrument Company, an SKF Group Company’s 
standard products division.  This machine is intended to be used for the detection of weak 
insulation within electric motors by trained professionals. It is intended to perform only the 
specified tests that this manual explains in detail. Please refer to chapters in this manual 
concerning specific operation of instrument. 
 

Содержание D12R

Страница 1: ...rby equipment Information furnished in this manual by Baker Instrument Company an SKF Group Company is believed to be accurate and reliable However Baker Instrument Company an SKF Group Company assume...

Страница 2: ...Buyer returns Products accessories or options to the Baker SKF factory for warranty service Baker SKF will bear the cost of return packaging and shipping and insurance on the return shipment For Produ...

Страница 3: ...ctions 13 On line labels 14 Safety precautions for setup 14 Initial tester power up and checkout 15 Using the footswitch 15 2 17 Test sequence voltages applicable standards 17 Recommended testing sequ...

Страница 4: ...tion Notes 44 Single phase motors and two terminal devices 44 Determination of a fault 44 Form coils 45 Determination of a fault 45 Notes and tips for form coils 45 Three phase motors 46 Determination...

Страница 5: ...nnection 69 Single phase lead connections 69 Note regarding 220 240 VAC units 69 Lifting the instrument 70 Supplement 2 71 PP185 power pack 71 Important safety information 71 Other Important Safety wa...

Страница 6: ...Bar to bar armataure testing accessory 95 Appendix A Winding Faults 97 Appendix B Troubleshooting 99 Self help and diagnostics 99 Step 1 Basic information 99 Step 2 Applications or service problem 99...

Страница 7: ...ury NOTICE NOTICE is the preferred signal word to address practices not related to personal injury Other Important Safety warnings Failure to follow these precautions could result in severe electrical...

Страница 8: ...pecifications 8 Ground the product This product is grounded through the grounding conductor of the power cord To avoid electrical shock plug the power cord into a properly wired grounded receptacle be...

Страница 9: ...tion requirements The unit is intended for use in Installation Category II Portable Equipment areas and pollution Degree II Environments where occasional non conducting condensing pollution can be enc...

Страница 10: ...s amended by Council Directive 93 68 EC and Council Directive 2004 108 EC on the approximation of the laws related to Member States relating to the electromagnetic compatibility as amended by Council...

Страница 11: ...Ray Tube CRT is the location where the tester displays test information At the Top menus corresponding to the four function keys above the CRT are shown The main portion of the screen displays the wa...

Страница 12: ...nterlock is disabled when this function is selected and the Footswitch is depressed AUX The auxiliary position is for use with a 30kV Power Pack The display will also show wave patterns from accessory...

Страница 13: ...rn 15 Leads energized indicator light This indicator light will illuminate when voltage is applied during a test V fast up When pressed this control will increase the applied voltage to a motor at an...

Страница 14: ...4 Seconds Div or micro Amps Div Setting Seconds division is displayed here during Surge Testing The micro amps division setting is displayed here during HiPot Testing Safety precautions for setup Ther...

Страница 15: ...y upon power up on and assures a positive ground is available to the tester If the instrument is not properly grounded the Open ground indicator will light and testing cannot proceed Check the supply...

Страница 16: ...Instrument Overview 2 24 2010 71 018 V11 EN D12R D6R D3R Users Manual 16...

Страница 17: ...ually low megohm value when compared to previous measurements or industry accepted limits for the type of insulation in the motor If a low megohm value is measured the motor should be inspected for gr...

Страница 18: ...vation for even doing the DA test is to reduce the test time to 3 minutes instead of 10 minutes for the PI test when the PI test may not be worthwhile To date there are no accepted values for the DA H...

Страница 19: ...1372 1029 575 469 1643 1232 600 490 1715 1286 2300 1878 6573 4930 4160 3397 11888 8916 6900 5634 19718 14789 13800 11268 39437 29578 IEC 34 15 V Line Per Unit 1 2 x 50 4E 5000 0 2us 65 480 392 6920 4...

Страница 20: ...chines IEEE 429 1972 Evaluation of Sealed Insulation Systems for AC Electric Machinery Employing Form Wound Stator Coils IEEE 432 1992 Guide for Insulation Maintenance for Rotating Electrical Machiner...