Aurora Rack Mount Manual Adendum

1

 

1. Aurora RM Features and Capabilities

1.1. Overview

Panametrics’ Aurora RM moisture analyzer makes it possible for a variety of industrial and laboratory gas processing 

facilities to check moisture content in real-time with high precision and reliability. The analyzer system is designed to 

continuously monitor moisture content in an extracted sample of process gas using tunable diode laser absorption 

spectroscopy (TDLAS). 
Similar to the Aurora H2O and Aurora Transport, it measures the partial pressure of water vapor (water in the gas 

state) in a gas sample introduced into its absorption cell.  Combined with simultaneous measurement of the sample 

pressure and temperature, it provides readings of user-selectable moisture parameters including dew point 

temperature, volume ratio and absolute humidity (lbs/MMSCF or mg/m

3

) and both analog and digital signal 

transmission in addition to temperature and pressure.

Figure 1: Aurora RM outside view showing rack-mounting rails and front panel.

It’s lightweight (<45 lbs) and rugged design with handles in the front and rack-mounting brackets pre-installed on 

the sides and the rear make it ideal for easy mounting in a standard 19” rack for quick deployment and use. The basic 

sample conditioning system allows for quick and easy gas connections on the rear panel, flow metering and 

adjustment on the front panel (see Figure 1), achieving stable readings within a few minutes of starting gas flow. 

Since the laser head is isolated from the sample gas, it provides non-contact drift-free moisture measurement 

tolerant to any contaminants such as particulates and liquid carryover in the sample

.

Analysis of the extracted sample uses TDLAS to determine moisture content in the gas with high accuracy and 

repeatability down to PPMv levels. The fast response of the Aurora RM quickly alerts the user when moisture 

concentrations are out of compliance. Once upsets are corrected and the process recovers, the fast response 

enables gas flow to be quickly cleared for entry into the process. 

1.2. Important Use Limitations

Environmental Conditions

The Aurora RM enclosure has an IP rating of 20, which implies the analyzer is designed to be used indoors and not 

designed to prevent dust or water ingress. The operator must be aware of weather conditions and coffee cup 

placement to prevent instrument damage or possible injury.

WARNING!

Liquids spilled onto the analyzer may leak into the electronics enclosure causing 

severe damage to laser control and measurement circuity, or possible user injury. 

Содержание Panametrics Aurora RM

Страница 1: ...H2O User manual P N BH003C11 rev C for operating the Aurora RM Please refer to the Aurora H2O User Manual for detailed instructions on operation general programming maintenance and troubleshooting in...

Страница 2: ......

Страница 3: ...material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective h...

Страница 4: ...ii no content intended for this page...

Страница 5: ...ervices including Training Product Repairs Service Agreements and more Please visit https www bakerhughes com panametrics panametrics services for more details CAUTION This symbol indicates a risk of...

Страница 6: ...at you bought has required the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemi...

Страница 7: ...Preface v no content intended for this page...

Страница 8: ...mounting in a standard 19 rack for quick deployment and use The basic sample conditioning system allows for quick and easy gas connections on the rear panel flow metering and adjustment on the front...

Страница 9: ...r horizontal flow cell is purged Dry down from ambient 5000 PPMv to 10 PPMv in seconds which enables very fast recovery from dehydration system upsets Direct readout in PPMv dew frost point C F densit...

Страница 10: ...Aurora Rack Mount Manual Adendum 3 2 Mounting the Aurora RM Rack Mount Analyzer Figure 2 Aurora Drawing for Rack Mounting P N 165M1371 rev A...

Страница 11: ...r user manuals for proper installation 3 Wiring the Aurora Rack Mount Configuration All the electrical connections input and output to the Aurora RM are made via the back panel as shown in Figure 3 be...

Страница 12: ...85 to If available make a third connection to the RS 485 network common to RTN Connect the shield wire from the cable to SHD terminal For operation on RS 232 connect RS 232 TXD to RS 232 RXD to and RS...

Страница 13: ...dous concentrations of flammable toxic or inert gases For safe and effective operation both the gas inlet and vent connections should be leak checked as ambient air leaks into the sample flow can lead...

Страница 14: ...sity of the laser beam or the accuracy of the moisture measurement When the Weak Signal error appears on the LCD display the laser absorption cell may need some maintenance Please contact Panametrics...

Страница 15: ...ewdriver and pry the fuse holder loose until your fingers can be used as shown in Figure 9 Gently pull on the fuse holder until it stops Swivel the fuse holder carefully as shown below and replace fus...

Страница 16: ......

Страница 17: ...eir respective holders BH005C11 EN D 09 23 Scan here for Panametrics support and service information Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S...

Отзывы: